You may also be interested in
CCD sensor, X-ray detector, Detector, Goniometer, Sensor
Text version of the page
mar 345 Image Plate DetectorSPECIFICATIONSmar dtb “desktop beamline” goniometer system.The Plate diameter345 mmUsable detector area93.480 mm At the forefront of modern X-ray structure work, the mar 345 Image Plate Detector is the perfect combina-tion of size, speed and efficiency featuring a huge diameter of 345 mm and read-out times as low as 26 seconds. It can be operated stand-alone or in combination with the mar 345 Image Plate Detector offers you the best choice for record- ing diffraction patterns of small molecules as well as protein crystals. Furthermore, it is a powerful tool for use in the fields of powder diffraction, texture analysis, small angle scattering, fiber and neutron diffraction.With its innovations in hardware design and software support the2 Diameter of scanned area180, 240, 300 or 345 mm (software selectable) Pixel size100 or 150 µm 2 (software selectable)Sensitivity1 X-ray photon per ADC-unit at 8 keV Energy range:4 keV to 100 keV X-ray photons Intrinsic noise: CMaximum humidity70 %Electricicty120 / 240 V (7.5 A)Energy consumption READ-OUT and CYCLE TIMESScanned diameter180 mm240 mm300 mm345 mmPixel size150 µm 100 µm150 µm100 µm150 µm100 µm150 µm100 µm Read-out time [sec]2634395356776896Total cycle time [sec]34424862668780108 |
|