Dynamometer Controller/Readout 6200
2Pages

{{requestButtons}}

Catalog excerpts

Dynamometer Controller/Readout 6200 - 1

M AGTROL 6200 Data Sheet Model 6200 Open-Loop Dynamometer Controller Features • • • • • • • • • • • • Open-Loop Dynamometer Control Built-in Pass/Fail Motor Testing Capability Interfaces: RS-232 and IEEE-488 High Speed Data Acquisition: 120 torque and speed points per second via IEEE bus (approx. 60/sec. via RS-232) High Quality, Easy-to-Read Vacuum Fluorescent Readout: Displays torque, speed, power and auxiliary values Current-Regulated Supply: Provides up to 1 amp output Adjustable Torque Units: English, Metric and SI are standard Dynamometer Overload Protection Internal Data Storage: Up to 100 data points Auxiliary ± 5 V DC Analog Input: For additional transducer Closed Box Calibration Rack Mounting: 19" (482.6 mm) with handles Description Magtrol's Model 6200 is an Open-Loop Controller designed for use with any Magtrol Hysteresis Dynamometer. The unit provides open-loop control of the dynamometer via an internal current-regulated power supply. With a high-quality vacuum fluorescent readout, the Model 6200 displays torque, speed and mechanical power values of the motor under test. In place of mechanical power, it can also display auxiliary transducer readings via the ± 5 V DC analog input. These displayed values can be stored internally or output via the RS-232 or IEEE-488 interface. Pass/Fail Motor Testing The Model 6200 comes with an easy-to-use motor testing Pass/Fail feature. This feature is ideal for quick pass/fail (go/no go) testing in production and incoming inspection applications. When the 6200 is operated in the Pass/Fail mode, one of three readings is used as the tested parameter: torque, speed or the auxiliary transducer. The two parameters not used are set with user-defined upper and lower acceptable limits. As the motor is loaded to the tested parameter value (for example, speed), the other two parameters (in this case, torque and transducer) are measured. Test results (for the other 2 parameters) are indicated with a “PASS” or “FAIL”, or the display can be toggled to show the actual values. System configuration Hysteresis Dynamometer Motor Under Test MODEL 6200 CONTROLLER CAUTION: DOUBLE POLE FUSING 75VA 50/60 Hz BRAKE ACCESSORY TORQUE-SPEED OUTPUT CTRL OUT DYNAMOMETER AUX. INPUT RS–232C GPIB/IEEE–488 EARTH GROUND BRAKE FUSE (5×20mm): UL/CSA 1.25A 250V SB IEC 1A 250V T MAGTROL, INC. BUFFALO, NY FUSE (5×20mm): 120V UL/CSA 800mA 250V SB 240V IEC 315mA 250V T Ordering Information 6200 Open Loop Dynamometer Controller 120 VAC 6200A Open Loop Dynamometer Controller 240 VAC www.magtrol.com

Open the catalog to page 1
Dynamometer Controller/Readout 6200 - 2

Specifications 6200 MEASUREMENT CHARACTERISTICS Maximum Torque 2000 units Maximum Speed 99,999 rpm Speed: 0.01% of reading from 10 rpm to 100,000 rpm Accuracy Torque: 0.2% of range (± 2 V) Aux: 0.1% of range (± 5 V) ELECTRICAL CHARACTERISTICS Brake: UL/CSA 1.25 A 250 V SB IEC 1.00 A 250 V T Fuses (5 × 20 mm) Power (120 V): UL/CSA 800 mA 250 V SB Power (240 V): IEC 315 mA 250 V T Power Requirements 75 VA Voltage Requirements 120/240 V 60/50 Hz Max. Compliance Voltage 45 V DC INPUTS AND OUTPUTS Auxiliary Input ± 5 V DC Accessory Torque: ± 2 V DC Torque/Speed Output Speed: 60 TTL pulses/rev,...

Open the catalog to page 2

All MAGTROL catalogs and technical brochures

  1. 4 QUADRANT

    2 Pages

  2. ZM Series

    5 Pages

  3. SK-01 Series

    3 Pages

  4. AMF Series

    3 Pages

  5. WB23 & WB27

    7 Pages

  6. Model 5200

    1 Pages

  7. 5211

    1 Pages

  8. 5251-2

    1 Pages

  9. SBB 14

    3 Pages

  10. FMF SERIES

    5 Pages

  11. ZS Series

    7 Pages

  12. RSB

    1 Pages

  13. PC-150

    2 Pages

  14. PCB

    1 Pages

  15. CST 113

    4 Pages

Archived catalogs

  1. WB/PB 65

    8 Pages

  2. WB/PB 15

    8 Pages