Application
Due to the extremely fast response time
of the patented M&C magneto-dynamic
measuring cell with no stagnant volume as
well as the negligible cross sensitivity from
other sample gas components, the M&C
oxygen-analyser PMA 100 has a wide variety
of applications. The analyser is a suitable and
reliable instrument for monitoring oxygen
concentrations in various gas analytical process
control applications including flue gas
measurements, inert gas installations, fermentation
processes and laboratory process
control installations.
Description
The M&C oxygen analyser PMA 100 is a
temperature controlled instrument which
has been designed for continuous measurements
of oxygen concentrations in particlefree
and dry sample gas. The PMA 100 is a
reliable and easy-to-operate instrument. It is
built into a housing for 19 inch rack mounting
/ table version.
The transducer unit maintains a constant
operating temperature of 55 °C.
The programmable measuring ranges are
displayed on the LCD display. Sample gas
connections as well as connectors for incoming
power supply and all output/input
signals are located at the rear side of the
analyser. The sample gas enters the analyser
via an external protective fine-filter. The
required flow rate can be adjusted at the
flow meter with needle valve, mounted on
the front panel upstream the M&C measuring
cell and the flow chamber with flow
alarm sensor and also with process pressure
sensor. The internal tubing is made of PTFE
and PVDF.
The PMA 100 has got nearly the same design
as the PMA 100-L. There are only a few differences:
the PMA100 is aditionnally equipped
with a pressure sensor, i.e. a pressure compensation.
Furthermore, there is an interface
type RS 232 and the possibility to drive 3
external solenoid valves for an automatic
calibration.
Oxygen Analyser Series PMAŽ
Version PMA 100
Microprocessor based,
19"-rack unit or table version
Special Features
Safe in operation
High accuracy and reliability
Physical measuring principle, linear
measuring ranges
Fast response time
Transducer unit temperature controlled
at +55 °C
Solid 19" metal housing
Large LCD- indicator with background
lighting
Selectable time constant for noise
damping if needed for unstable process
conditions
4 free selectable O2 measuring ranges,
including zero suppression, lowest span
1%
Manual, automatic or remote range
control
User friendly menu control corresponding
to NAMUR recommendations
Isolated output signal 0/4-20mA
4 configurable MIN. or MAX. oxygen level
alarms
4 free configurable isolated binary inputs,
for example: alarm confirmation, measuring
range control, autocalibration start
Process pressure measuring and
compensation
Analogue signal memory during
adjusting and calibration
Manual / automatic calibration
3 code levels, access to 2 code levels only
with release code
Interface RS 232 type
Flow alarm monitoring
4 free configurable potential free relay
contact outputs, for example: oxygen
level alarm
option: interface RS 485 type
Measuring principle of M&C
oxygen analyser
The PMA 100 utilises the paramagnetic
principle of operation to measure oxygen
concentration. The analyser measures the
paramagnetic susceptibility of the oxygen
in the sample gas by means of the M&C
magneto-dynamic measuring cell. The physical
propertys which distinguishes oxygen
from other gases is its paramagnetism. It
is significantly higher comparing to other
common gases. This operation principle is
one of the most accurate and reliable procedures
to determine the oxygen concentration
in a gas mixture from 0 to 100 Vol.%.
The robust M&C cross-flow cell has no stagnant
volume. Advantages are the fast
response time, the flow rate up to 60 l/hr,
the small volume of 2 ml, the extremely low
drift, the absolute linearity and the negligible
cross sensitivity against other sample gas
components. With a proper sample conditioning
and pressure, the M&C cell will never
need replacing.
PMA 100
|
|