Text version of the page
MMSPFA+ Series Pelton Wheel flowmeter with 316 stainless steel body, providing a pulse output. Model Type Flow Rate Range Pressure Drop Pipe Size Weight MMSPFA.12.5 1 to 65 l/min 0.2 bar @ 32l/min; 0.9 bar @ 65 l/min ½” BSP female 2.5 kg MMSPFA.20 2 to 130 l/min 0.17 bar @ 65 l/min; 0.69 bar @ 130 l/min ¾” BSP female 2.5 kg MMSPFA.25 3 to 160 l/min 0.17 bar @ 80 l/min; 0.69 bar @ 160 l/min 1” BSP female 4 kg MMSPFA.30 5 to 220 l/min 0.17 bar @ 110 l/min; 0.69 bar @ 220 l/min 1¼” BSP female 4 kg MMSPFA.37 7 to 350 l/min 0.17 bar @ 175l/min; 0.7 bar @ 350 l/min 1 ½” Wafer fitting 4 kg MMSPFA.50 10 to 500 l/min 0.17 bar @ 250 l/min; 0.7 bar @ 500 l/min 2” Wafer fitting 4 kg MMSPFA.62* 20 to 800 l/min 0.17 bar @ 400 l/min; 0.7 bar @ 800 l/min 2 ½” Wafer fitting kg MMSPFA.75* 25 to 1300 l/min 0.17 bar @ 650 l/min; 0.7 bar @ 1300 l/min 3” Wafer fitting kg MMSPFA.100* 25 to 2200 l/min 0.17 bar @ 1100l/min; 0.7 bar @ 2200 l/min 4” Wafer fitting kg MMSPFA.150* 100 to 5000 l/min 0.17 bar @ 2500 l/min; 0.7 bar @ 5000 l/min 6” Wafer fitting kg MMSPFA.200L* 300 to 5500 l/min 0.17 bar @ 4000 l/min; 0.7 bar @ 8000 l/min 8” Wafer fitting kg *Lower flows on request Common Features: Pressure Rating 40 bar [120 bar option, add .120 to model type] Temperature Rating 70° C Viscosity Range Up to 10 cPs; ring factory if higher Accuracy ±2% of actual reading for the top 90% of range. ±0.5% FSD for the bottom 10% of range. [003 and 005 linearisation required] Repeatability ±0.2% reading over top 90% of range Body and Internals 316SS with Teflon PFA rotor and sapphire/sapphire bearings Ferrites Sealed Seal FPM ‘O’ ring (subject to chemical compatibility). [120 bar version has PTFE gasket] Wetted parts of sensing electronics 316SS External materials of sensing electronics 316SS or 303SS and anodised aluminium Output & Power NPN open collector with 4-way M12 sensor connector [IEC-61076-2-101] – e.g. compatible with Brad Harrison, Hirschmann, ERNI and Turck Banner. Supplied with 5m cable – bare ends. 8 to 28Vdc input. Options: Output PNP or TTL (identified by green or blue cap respectively). Also divide by 2, 4 or 10 function Connections Plain OD pipe for butt welding or flanges available in alternate sizes and styles Flow rate, temperature and pressure Alternate ratings available in different sizes and styles, consult factory for further details Seals FEP covered Silicon, EPDM or Kalrez® for chemical protection IS Hazardous Area Version ATEX Available in alternate styles to EEx ia IIC T5 – consult factory for further details (limited to process temperatures up to 70°C) Consult factory for input and output data. Housed in an IP65 epoxy coated die-cast aluminium body-mounted enclosure with 5pin DIN socket. SB or BB IS Electronics pickup For use either when instrumentation is also included in the hazardous area or an M20 gland is required. Body-mounted stainless steel or Naval Brass IP68 enclosure, one or two M20x1.5 gland entries. Documentation: Operating and Maintenance Manual LM0111 with General Information LM0092 Installation Drawing C6089 or C3678 and general installation C3729 Conformity: These products conform to PED and EMC. ATEX option available. Flowmeter Specification Sheet

pageCatalog pdf di En 2012-05-22-30