TARGA III Dynamic Runout System
The TARGA III stands alone as the only dedicated, capacitive-based system to
monitor dynamic spindle runout.
High-Resolution Capacitive Sensing
High reliability, resolution, and accuracy
Target Selection
Select gage pin sizes of 1/8", 2mm, 1.75mm
Metric and Inch Units
Display with the measurement units you want.
ISO 9000 Compliant
Calibration is NIST traceable
Manufactured by Lion Precision
For over thirty years, we have been a world leader in the design of high-resolution, high-speed capacitive gaging. Our customer service is second to none.
No Export License Required
ECCN Classification: EAR99
Dynamic Runout
Dynamic Runout, now industry standard terminology, was first coined by Lion Precision when the original TARGA Dynamic Runout System was introduced. It was the first and only system that could measure spindle runout at operating speeds. The only runout measurement that really counts.
The TARGA III does not measure RPM or provide printed reports.
Specifications
BNC Output Voltage
0-10VDC - TIR, DRO Modes ±5VDC - Other modes
BNC Output Voltage Scaling
Meter units: 25jum/V Inch units: 0.001"/ V
BNC Output Resolution
625nm, 0.000,025"
Display Resolution
0.5jum, 0.000,02"
Measurement Range
250jum, 0.010" (1/8" pin)
Near Gap
125jum, 0.005" (1/8" pin)
System Power In
±15VDC @ 0.25A
External Power Supply (Included)
Input:
100-240VAC, 50/60Hz
Capacitive vs. Eddy-Current Sensors
Some runout sensors use eddy-current technology. Eddy current technology does not work well with rotating, ferromagnetic targets.
The Targa III uses capacitive sensors which are not affected by a rotating target. This chart shows runout measurements of the same spindle using eddy current and capacitive sensors.
Eddy-Current Sensor
Capacitive Sensor
563 Shoreview Park Rd. St. Paul, MN 55126 651-484-6544 www.lionprecision.com