| | | • Very simple operation, easily-read display (backlighting selectable) • Changeable reference sensors (plug & play) this means, up to 10 pressure ranges with one instrument • Ranges from 0...5.8 psi (400 mbar) up to 0...58000 psi (4000 bar) (also vacuum-, compound- and absolute pressure ranges) • Accuracy up to ±0.025% f.s. resp. 0.1% of reading • Logging function, pressure rate, min-/max value, Bargraph, Tare • Issue calibration certificates with optional PC-Windows®-software LPC-Cal • Chargeable Lithium-Ion battery, incl. Charger, USB-interface • Switchable pressure unit PSI, bar, mbar, kPa, mmHg, inHg Together with a suitable calibration pressure source like the calibration hand pumps LPP 30 (pneumatic) or LPP 700 / LPP 1000 (hydraulic), the LPC 200 offers a pecision pressure calibration system, suitable also for use on site, portable. The LPC 200 also opens up other helpful applications, e.g. recording of pressure curves (logging function), leak testing (pressure rate function). These useful functions can be easily activated or deactivated by pressing the respective keys. A factory calibration certificate for each reference sensor certifies the total uncertainty of the measuring chain. Alternatively, on request, a DKD calibration certificate can be | | |
| | | Using the optional LPC-Cal software, the logged data, stored in the LPC 200, can be transferred to a PC via a USB connection, where it can be stored in an Excel® file for documentation and further evaluation. In addition, LPC-Cal offers the possibility of PC-supported calibration of pressure measuring instruments using the LPC 200. | | |
| | | | | | | | | | | Technical Data (Indicator LPC 200 together with reference sensor LPC-2): | | | | Standard Pressure Ranges (PSI) Standard Pressure Ranges (bar) | 5.8 23.2 87 0.4 1.6 6 | 232 580 1450 3626 8700 14500 16 40 100 250 600 1000 | | | | Overpressure safety (bar) | 2 10 35 | 80 80 200 500 1200 1500 | | | | Burst pressure (bar) | 2.4 12 42 | 96 400 800 1200 2400 3000 | | | | Accuracy (Total uncertainty of measurement chain) | 0.025% FS or 0.1% of measured value (whichever is the greater), calibrated at 23°C and in vertical mounting position with pressure conn. facing down. | | | | High Pressure Ranges (PSI) High Pressure Ranges (bar) | 23200 36200 58000 1600 2500 4000 | | | | | Overpressure safety (bar) | 2000 3000 4400 | | | | | Burst pressure (bar) | 4000 5000 7000 | | | | | Accuracy (Total uncertainty of measurement chain) | 0.15% FS (calibrated at 23°C and in vertical mounting position with pressure conn. facing down.) | | | | Type of pressure | see above (standard and high pressure ranges) but also vacuum-, compound- and absolute pressure ranges | | | | Units of pressure | bar, mbar, kPa, mmHg, inHg, PSI (depending on range, switchable) | | | | Active temperature compensation | +10...+40°C | | | | | Permissible ambient temperature | 0...+50°C | | | | | | | | | | | |