Catalogue Digital Test Pressure Gauge LDM 70
www.druck-temperatur.de
print switch display
Page / 2
Contact the
Manufacturer
Request
a Quote
LEITENBERGER - 144257, 148039, 2189, 14994
/ 2
See other catalogues for LEITENBERGER
Text version of the page
LR-Ca/
Digital Test Pressure Gauge LDM 70 battery operated, accuracy 0.25% or 0.5% f.s.
LDM 70
Digital Reference Test Pressure Gauge LDM 70 LDM 70-E25: Accuracy ±0.25% f.s. / LDM 70-K50: Accuracy ± 0.5% f.s.
• Battery powered, rugged, rotatable housing (DS 70 mm)
• 4.5-digit LC display (digit height 11 mm)
• 6-digit additional display to show the selected pressure unit (height 7.5 mm)
• Pressure unit switchable (PSI-bar-mbar-mWs-inHg-cmHg-mmHg-kPa-MPa)
• Easy configuration via three push buttons
• Ingress protection IP 65
• Ranges from 0/1.45 psi to 0/8700 psi ( 0/100 mbar to 0/600 bar)
• MIN and MAX value storage
The display is continuously rotatable so that a clear readability is guaranteed even in unusual installation positions. The LDM 70 can be operated menu-driven via three pushbuttons. Besides showing information about the nominal pressure range as well as minimal and maximal pressure of the process, several pressure units (PSI, bar, mbar, mWC, inHg, cmHg, mmHg, kPa, MPa) and the position of the decimal point can be set. The instruments zero and span can be recalibrated, the LDM 70 can be set to factory default values via menu.
Technica data:
Type LDM 70-E25
Type LDM 70-K50
Pressure Ranges: (see tab. on back page)
all standard ranges, gauge pressure, absolute pressure, vacuum and compound ranges
all standard ranges 0/1.6 bar to 0/400 bar
Nominal case diameter:
DS 70 mm / 27.6 inches
Accuracy (as per IEC 60770):
±0.25% f.s (ranges < 5.8 psi / 400 mbar: ±0.5% f.s.)
±0.5% f.s.
Measuring rate:
5 per second
Thermal error
(zero-point and span within temperature compensated range)
Tolerance band:
Ranges < 0.1 bar: <± 2.0% f.s.; average TC: ±0.3% f. s. / 10 K Ranges < 0.25 bar: <±1.5% f.s.; average TC: ±0.2% f.s. / 10 K Ranges < 0.4 bar: <1.0% f.s.; average TC: ±0.14% f.s. / 10 K Ranges <1.0 bar: <0.1% f.s.; average TC: ±0.1% f.s. / 10 K Ranges >1.0 bar: <±0.75% f.s.; average TC: ±0.07% f.s. / 10 K
<±0.2% f.s. / 10 K
Temp. compensated range:
Ranges <0.4 bar: 0...50°C; ranges >0,4 bar: 0...70°C
-25...+85°C
Permissible temperatures:
Medium: -4/+185°F=-20/85°C; Ambient: -4/+158°F=-20/70°C; Storage: -22/+176°F=-30/80°C
Material display housing:
PA 6.6 polycarbonate
Material pressure port:
stainless steel 1.4571 / 1.4435
stainless steel 1.4435
ceramic Al2O3 96%
Material seals:
wetted seals: FKM (optional: EPDM, NBR, others on request)
Display:
LCD 1.57" x 1.18"= 40 x 30 mm; 4 1/2 digit 7-segment-main display, digit height 11 mm, Indication range ±19999; 6-digit 14-segment additional display, digit height 7.5 mm
Electromagnetic compatibility:
emission and immunity according to EN 61326
Mechanical stability:
vibration: 5 g RMS (20...2000 Hz); shock: 100 g / 11 msec.
Supply:
3.6 V lithium batteries, 2 pcs. (type 1/2 AA)
Data memory:
EEPROM (non-volatile)
Ingress protection:
IP 65
Installation position:
any (the LDM 70 is calibrated in vertical position with pressure port down. If this position is changed on installation, there can be slight deviations in the zero-point for LDM-70-E25 ranges <1 bar
Other data:
weight 300 g; AD-converter solution: 10 Bit;mech. operational life: > 100 x 106 pressure cycles; battery life: app.. 5 years in standby mode;
Pressure port:
1/2" BSP male, 1/4" BSP male, 1/2" NPT male, 1/4" NPT male
Die Digital test pressure gauges LDM 70 are available as complete pressure calibration kits (case, with calibration handpump LPP 30). The calibration handpump LPP 30 generates pressure up to 500 psi (35 bar), switchable to vacuum generation -28 in Hg.
For higher pressure ranges, the Digital test pressure gauges LDM 70 can be supplied as pressure calibration kit together with the hydraulic pressure comparators LSP series.
DRUCK & TEMPERATUR Leitenberger GmbH
Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • Germany
Tel.: +49 - 71 21 - 9 09 20 - 0 • Fax: +49 - 71 21 - 9 09 20 - 99
E-Mail: DT-Export@Leitenberger.de • http://www.druck-temperatur.de
Technische Änderungen vorbehalten. Freibleibend • (Rel. 070306) • All technical modifications reserved. Without engagement
DirectIndustry's Virtual Technical Library: PDF Catalogue | Technical Documentation | Brochure | Manual | Industrial directory | Specifications | Characteristics
Search Go
page 1 p.1
page 2 p.2
pdf-page pdf di En 2008-12-49-05