Deadweight Tester / Pressure BalancePneumatic and Hydraulic
CPB5000
Dimensions in mm
Adapter for specimen connection
Manual quick-connection clamps with threaded adapters Available Pressure Ranges: Pneumatic Models Hydraulic Models >
suitable for dry and clean air or nitrogen(for other media available on request) suitable for operating fluid CPB5000-FLUID (1 L included) (for other media available on request) Range [bar]Range [psi]Range [bar]Range [psi]-0.03 ... -1 (Vacuum)-0.435 ... -14 (Vacuum)0.2 ... 600.03 ... 20.435 ... 300.2 ... 1002.9 ... 10000.2 ... 102.9 ... 1501 ... 25014.5 ... 50000.4 ... 505.8 ... 5001 ... 40029 ... 100000.4 ... 1005.8 ... 10002 ... 60029 ... 150005.8 ... 15002 ... 1000 Accuracy (uncertainty): Two versions available: ±0.015% or ±0.010% of measured value. The accuracy ischaracterised by the deviation span, which is the sum of the systematic error and the uncertainties of measurement.Longterm stability is not taken into account. Please note that corrections might be required if the instrument is usedwothout the optional Intelligent Calibration Module IKM - see operating manual. Other Models: For high pressure up to 4000 bar / 60000 psi: see datasheet CPB5000-HP For differential
pressure calibration: see datasheet CPB5000-DP >
DRUCK & TEMPERATUR Leitenberger GmbH Bahnhofstr. 33 D-72138 Kirchentellinsfurt GermanyTel.: +49 - 71 21 - 9 09 20 - 0 Fax: +49 - 71 21 - 9 09 20 - 99E-Mail: DT-Export@Leitenberger.de http://www.druck-temperatur.de
Technische Änderungen vorbehalten. Freibleibend (Rel. 070618) All technical modifications reserved. Without engagement