Calibration Test Pump LPP 30 (pneumatic)Pressure generation 28 in Hg...+500 psi
LPP 30
Calibration Test Pump LPP 30, pneumatic operatedto generate pressure up to +35 bar / +500 PSI
switcheable to generate vacuum up to -0.95 bar / -28 inHg
LPP 30 , the samepressure / vacuum is applied to the two instruments when the pump
is operated.Despite its compact dimensions, the calibration test pump LPP 30 iseasy to operate and allows for exact generation of the required test pressures. A change-overswitch enables the generation of vacuum as well. The LPP 30 is fitted with a fine adjustmentvalve for the precise adjustment of pressures. The reference instrument is screwed directly on to
the top of the pump and the unit under test is connected by means of a connection tube
incorporating an adapter 1/4" BSP female (or 1/4" NPT female) port. The Calibration Test Pump Specification: LPP 30 is used to generatepressures and vacuum for checking, adjusting and
calibrating mechanical and electronic pressure measuring instruments by comparative measurements.
These pressure tests may be carried out in laboratories,
workshop or on site at the measuring point. If the
instrument under test and a sufficiently accurate reference
measuring instrument (e.g. Electronic Pressure CalibratorLPC 300, ±0.025% FS) are connected to the
Pressure generation:0...35 bar = 0...500 PSI (or equivalent), switchable to
Vacuum generation:0... 0,95 bar = 0...-28 inHG (or equivalent)Pressure ports:1/2" BSP female rotating (with
industrial seal) for reference instrument on the top of the pump(Type LPP-30-N14: 1/4" NPT female)1/4" BSP female for unit under test (at connection tube)(Type LPP-30-N14: 1/4" NPT female)Materials:anodized aluminum, brass, ABS
Fine adjustment:Fine adjustment valveDimension:app. 220 x 105 x 63 mmWeight:app. 510 g >
Included in standard supply:
•Calibration test pump LPP 30 or LPP 30-N14 •Connection tube with 1/4" BSP female rotating port (with O-ring)or with 1/4" NPT female port at type LPP 30-N14•operating manual >
Optional accessories:
•Set of adapters and seals for unit under test•Rigid plastic case with
industrial foams >
Recommended Reference Instruments:
Calibration Test Pump LPP 30 with Electronic calibrator LPC 300 •Accuracy ±0.025% FS:Electronic
Pressure calibrator LPC 300 (see sep. datasheet)•Accuracy ±0.050% FS:Electronic Pressure Calibrator TLDMM (see sep. datasheet)•Accuracy ±0.2% FS:Digital Test
Pressure gauge TLDMM-A02 (see sep. datasheet)•Accuracy ±0.2% FS:Digital Test Pressure Gauge TDMC (see sep. datasheet)•Accuracy ±0.5% FS:Digital Test Pressure Gauge DTG 35 (see sep. datasheet)•Accuracy ±1.0% FS:Analogue Pressure Gauge LPP-MANO (see back page) >
DRUCK & TEMPERATUR Leitenberger GmbH Bahnhofstr. 33 D-72138 Kirchentellinsfurt GermanyTel.: +49-7121-90920-0 Fax: +49-7121-90920-99E-Mail: DT-Export@Leitenberger.de http://www.druck-temperatur.de
Technische Änderungen vorbehalten. Freibleibend (Rel. 060519) • All technical modifications reserved. Without engagement