Analex Brochure - Kittiwake - #6

/ 6


catalogue search
P. 01
P. 02
P. 03
P. 04
P. 05
P. 06
Pages:


See other catalogues for Kittiwake

Text version of the page
ANALEXpq
PQ® Technology Ferrous Debris Monitors
On-screen instructions provide a step-by-step guide to using the monitor and a printed user manual includes practical tips gathered from users worldwide. CALIBRATION
A standard 750' calibration block is supplied with every ANALEXpq unit, to ensure consistent and reliable measurements. This block is measured during the calibration routine and the units' internal calibration data is adjusted to a PQ Index of 750. Recalibration is required only when the unit gives consistent readings on the 750' standard which are not within the +/-1% permitted range.
ANALEXpq Ferrous Debris Monitors are constructed using sophisticated dual-coil magnetometers for greater measurement accuracy.
When no oil sample is present, the sample coil (sensor) and the reference coil are in balance. Both coils are designed to maintain this balance in conjunction with changes in the ambient temperature.
A sample of oil containing ferromagnetic debris is placed on the sensor, thus altering the balance between the coils. The resulting out-of-balanee' signal is amplified, filtered and displayed as a PQ® Index, which relates directly to the mass of the ferromagnetic debris in the sample.
SPECIFICATIONS
ANALEXpqM
ANALEXpqA
AC power input
115/230V ± 10%
115/230V ± 10%
Supply frequency range
47 to 63 Hz
47 to 63 Hz
Power consumption
75 watts
15 watts
Ambient temperature
20°C - 30°C
20°C - 30°C
Electrical connection
IEC 6A 3-pin connector
IEC 6A 3-pin connector
Serial connector
9-pin D (male plug required)
9-pin D (male plug required)
Display
4 line 76 x 25 mm back lit alphanumeric LCD
2 line 11 mm back lit alphanumeric LCD
Dimensions
360 x 360 x 140 mm
560 x 205 x 470 mm excl. sample stacker or sample tray
Weight
5.5 kgs
12 kgs
Maximum sample diameter
53 mm aperture of measurement carousel
53 mm aperture of measurement carousel
Repeatabilty
± 1% relative or 4 PQ* Index units
± 1% relative or 4 PQ* Index units
Kittiwakel\
Measuring, Monitoring & Managing
Kittiwake Developments Ltd. ^
3-6 Thorgate Road, Littlehampton, West Sussex BN17 7LU, United Kingdom Tel: +44 (0) 1903 731 470 Fax: +44 (0) 1903 731 480 Email: sales@kittiwake.com Web: www.kittiwake.com