| | | | | |
| | | |
| | Digital Manometer with Switch Outputs | |
| | | |
| | USER SELECTED SWITCHING POINTS AND FUNCTIONS | |
| | | |
| | The dV-2 PS is a sophisticated digital pressure gauge with two independent pressure switch outputs. The gauge employs KELLER'S proven microprocessor technology which produces highly accurate results. All readings are linearised and temperature corrected, to give an accuracy of 0,2 %FS. There are two independent switches for High and Low control functions. These are solid state and therefore not subject to wear. The switch points are programmed directly from the two front panel buttons and a menu system. Alternatively, the dV-2 PS can be connected via the RS485 link to a laptop/PC, and configured using KELLER'S "Pressure Switch Console" software. User defined configurations can be stored on the computer for future reference. Pressure sequences can also be monitored and recorded on the PC for application diagnostic purposes using KELLER'S Read30 software. | |
| | | |
| | Two independent floating switches are available; neither is subject to wear and they can be operated individually as make or break contacts. The switch can provide a window function or be utilised as a hysteresis switch. These switching functions enable the dV-2 PS to assume control ' dV-2 PS tasks. Respective switch output status is displayed on an LED located on the front. | |
| | | |
| | dV-2 PS has the following functions: | |
| | | |
| | HIGH/LOW The maximum and minimum values are indicated via the Select button. RESET Allows the max. and min. value to be reset at the actual pressure value. ZERO Sets the zero to any value as a new zero reference. Barometric pressure vari- ations can thus be compensated. The factory setting of the zero for the ranges < 31 bar is at 0 bar absolute. For sealed gauge pressure measurements, activate "ZERO SET" at ambient pressure. Instruments with ranges > 31 bar are calibrated in a sealed gauge mode with ambient pressure as a zero reference. SWITCH Configuration of the switch outputs. Switch points, switch window, switch hys- teresis, open and close. | | |
| | | |
| | | The pressure can be displayed in bar, MPa or PSI. | |
| | | |
| | | | | | | | | | | | | | Pressure Ranges absolute " relative Display Resolution | 0...4 bar 0...31 bar 0...200 bar 0...700 bar -1...3 bar -1...30 bar 1 mbar 10 mbar 100 mbar 100 mbar | | | | Overpressure Accuracy * Temperature Range | 1,5 x pressure range 0,1 %FS 10...40 °C | | | | Measuring Interval / Switch Delay Supply Switch Outputs Smallest Resolution of Switch Points Pressure Connection Electrical Connection Interface Protection Width x Height x Depth Weight | ~ 30 ms 8...28 VDC 2 pes. PhotoMOS-Relals 40 V / 0,4 A Same as display resolution G 1/4" or Cutting ring/clamping ring, 0 6 mm (only up to 600 bar). Optional: other connections. Binder (8 pole) or PG-7 cable Input with spring clip In the Instrument RS 485 IP 64 68 x 70 x 35 mm ~ 100 g (with rubber protection ~ 136 g) | | | | | - Protective rubber covering, front display with customer logo | | | | * Includes linearity, repeatability, hysteresis and temperature error. Does not Include Zero stability. | | | | | | | | | KELLER AG fur Druckmesstechnlk KELLER Ges. fur Druckmesstechnlk mbH | St. Gallerstrasse 119 CH-8404 Wlnterthur Tel. +41 (0)52 - 235 25 25 Fax +41 (0)52 - 235 25 00 Schwarzwaldstrasse 17 D-79798 Jestetten Tel. +49 (0)7745 - 9214 - 0 Fax +49 (0)7745 - 9214 - 60 | | | | | | | | |
| | | |
| | Companies approved to iso 9001 / en 29001 | | |
| | | |
| | | | | |