Electrolytic Conductivity
JUMO BlackLine Lf-GT/-EC2-electrode conductivity cells JUMO tecLine Lf-VA 2-electrode conductivity cells
Compact conductivity probes with an outstanding
price/performance ratio.JUMO BlackLine Lf-GT,
industrial graphite electrodes cell constant K = 1.0 and 3.0 for ranges up to 200mS/cm attached cable or screw-in version with Pg13,5 120mm fitting length, 12mm diameter Rugged versions in stainless steel or titanium, with a wide variety of process connections. usable up to 90°C and pressure up to 6bar for measuring ranges from 0.05µS/cm to 1 mS/cm integrated
temperature probe (option) pharmaceutical version with factory certificate to EN 10204-3.1 and FDA verification JUMO BlackLine Lf-EC, electrodes in titanium or stainless steel 1.4571 high-temperature version up to 200°C and 17bar cell constants K =0,01, 0,1 and 1,0 G1/2" screw-in thread with attached cable or ½"NPT from 40mm fitting length Possible application areas Pure and high-purity water, pharmaceuticals, chemicals,
food technology. chip production, ion exchangers, reverse
osmosis plant. usable up to 100°C and pressure up to 6bar integrated Pt100 temperature probe as immersion version with immersion length up to 1500mm (option) Further information: Data Sheet 20.2924 Possible application areas >
JUMO tecLine Lf-GT2-electrode conductivity cells
Drinking and surface water monitoring, ion exchangers andreverse osmosis plant, HVAC and
refrigerating systems, horticulture, seawater and freshwater aquariums, lightly
contaminated industrial process and flushing water. Further information: Data Sheet 20.2922 >
JUMO ecoLine Lf-PVC2-electrode conductivity cells
Industrial version with graphite electrodes for measuring ranges up to 200mS/cm. cell constants K=1,0, 3,0 or 10,0 various process connections provide optimum adaptation to process conditions Possible application areas Separating media, drinking water treatment, wastewater
checks, concentration monitoring, service water treatment. Further information: Data Sheet 20.2925 >
JUMO tecLine Lf-TA2-electrode conductivity cells
Tried and tested versions for industrial use. measuring ranges: 0 – 1 mS/cm (K = 0.1) or 0.01 – 15mS/cm (K = 1.0) temperature range up to 55°C and pressures up to 6bar at 20°C Well-established versions for industrial use. stainless steel version, K = 0.1, with three metallic pin electrodes, graphite version K = 1.0 typical measuring ranges: 0 – 1 mS/cm (K = 0.1) or 0.01 – 15mS/cm (K = 1.0) PVC flow-through fitting can be supplied temperature range up to 90°C, and pressures up to 6bar at 20°C Possible application areas Possible application areas General water engineering,
industrial refrigeration and air condi-
tioning systems, drinking and bathing water. Immersion model up to 2000mm for containers and open