MAXItube™ 160kV
MAXItube™ 1
The MAXItube™ 160kV is a 160 kV,
water cooled stationary anode metal
ceramic x-ray source.
This source is specifi cally designed
for Non-Destructive Imaging Applications.
MAXItube™ SPECIFICATIONS
Maximum Peak Voltage 160 kV
Focal Spot EN12543
Small D = 1.0 mm
Large D = 3.0 mm
Focal Spot IEC
Small 0.4 mm
Large 1.5 mm
Target Angle 20°
Maximum Continuous Rating
Small 640 Wwith 4 Litre/min cooling fl ow
Large 1600 Wwith 4 Litre/min cooling fl ow
Cooling Medium Water
Reference Axis Perpendicular to port face
Radiation Coverage 40°
Loading Factors for Leakage Radiation 160 kV, 10 mA
X-Ray Tube Assembly Permanent Filtration 0.8 mm Be
High Voltage Cable R24
Weight (approx.) 8.0 kg ( 17.6 lbs)
OPERATION
GENERAL
The control of the high voltage and the
fi lament current as well as the design of
the cooling unit is the responsibility of
the equipment manufacturer.
High Voltage Connection
One terminal type R24 high voltage cables
supply high voltage and fi lament
current.
Two diff erent methods of mounting the
high voltage cable are available.
Cooling of the Anode
It is the responsibility of the customer
to ensure that the cooling medium fl ow
meets the required cooling conditions.
Insuffi cient cooling of the anode can lead
to the destruction of the anode, therefore
cooling must be switched on before
the application of high voltage.
Control of Cooling Medium
Flow, pressure, and temperature of the
cooling medium at the inlet to the tube
or the tube assembly must be appropriately
monitored. High voltage must be terminated
when the pressure or fl ow rate
fall below the minimum level or when
the temperature exceeds the maximum
level. When the tube is switched off the
coolant fl ow must continue for at least
2 minutes in order to protect the anode
from destruction.
|
|