Datenblatt TIDAS MMS-VIS/NIR VIS/NIR spectrometer (300nm - 1100nm) - 1 Pages

  1. P. 1

Catalogue excerpts

J&M Analytik AG Willy-Messerschmitt-Straße 8 73457 Essingen Datasheet spectrometer MMS VIS/NIR 3011 Type Module Connector Entrance slit with cross section converter Grating PDA Hamamatsu S3904 Pixels Wavelength range Pixel distance Opt. Resolution according to USP Wavelength accuracy Repeat accuracy Wavelength temperature drift Scattered light Dimensions [LxDxH) Baseline drift S/N ratio *1) *2) Unit Value µm MMS-VIS/NIR FSMA 905 70 x 2500 L/mm 256Q nm nm/pixel nm 256 300 … 1100 ~3,3 < 10 nm nm nm/° C < 0.5 < 0.1 < 0.005 % mm³ AU/h AU <0,8 @450nm ~ 67* 58nn * 40 5*10E-04*1) < 3*10E-05*2) Baseline drift will be measured at 254 nm after 10h warm up @ 21° C±2° ambient C temperature according to ASTM E685 Noise will be measured at 254 nm after 10h warm up @ 21° C±2° ambient C temperature according to ASTM E685, without methanol flow with the following settings: • • • • Integration time <100ms Pixel bunching 2 (2x2,2nm ~ 4nm) Integration time x accumulation <2 sec. Detector saturation ~ 80%,

 Open the catalogue to page 1

All J&M Analytische Mess- und Regeltechnik catalogues and technical brochures

  1. TIDAS S MSP 800

    2 Pages

    En

Archived catalogues

  1. TIDAS 100 / 116

    2 Pages

    En