Part number TVOCV1.6 This publication is not intended to form the basis of a contract and specifications can change without notice.
TVOC is a fixed photoionisation detector (PID) designed for continuous
measurement of total volatile organic compounds (VOCs). New ATEX
approvals enable a 3 wire TVOC system to be used in Zone 2 hazardous
areas, without safety barriers. ATEX and IECEx approvals allow TVOC to
be used in Zone 1 hazardous areas, with safety barriers.
The 4-20 mA analogue output enables TVOC to be integrated into a DCS
control system to give warning or control of high VOC levels in the
working environment.
TVOC utilises a diffusive sample technique resulting in less
contamination issues compared to pumped systems, reducing lamp
cleaning and servicing requirements. Simple to install, service and
calibrate, TVOC requires no hot work permit and the PID sensor is
accessible and changeable in a matter of seconds.
TVOC SystemPlus comes complete
with a pumped sample system for
applications where a sample needs
to be drawn to the unit from as far
away as 30 meters. SystemPlus
incorporates a TVOC unit together
with pump, sample preconditioning
with flow cut off switch, optional USB data
logger and power supply all contained in a
waterproof housing.
181mm (7.1”)
125mm (4.9”)
125mm (4.9”)
Applications include:
• Manufacturing • Processing • Refineries and petrochemical
• Offshore • Chemical • Waste water treatment • Pharmaceutical
• Indoor air quality • Pulp and paper • Solvent recovery systems
• Industrial painting and coating
TECHNICAL SPECIFICATION
SENSOR (PID)
10.6 eV lamp with Ion Science patented Fence Electrode
technology*
HAZARDOUS AREA SAFETY APPROVALS
ATEX: II 3 G Ex nA IIC T4 (-20°C Ta 50°C)
IECEx: II 2 G Ex ia IIC T4 (-20°C Ta 50°C)
INGRESS PROTECTION RATING
Designed to - IP65
Dependent on cable glands fitted Sensor - IP53
POWER
5-28 VDC Max 65 mA
OUTPUT
4-20 mA (requires a 8-35 VDC power supply)
For IS requirements 8-30 VDC power supply
RANGE
0-10 ppm, 0-100 ppm, 0-1000 ppm User Selectable
SAMPLING
Diffusion (can be pumped if required using an adaptor)
DISPLAY
7 segment, 4 digit LCD, 4 color LEDs
RESPONSE
Sensor - T90 25%
CALIBRATION
Magnetically accessed 100 ppm Isobutylene via calibration
kit accessory
TEMPERATURE
Operating: -20 to 50 oC (-4 - 122 oF) Humidity: 0-95% RH
(non- condensing)
WEIGHT & DIMENSIONS
Instrument: 1.4 kg (3.1 lb) Packed: 1.6 kg (3.5 lb)
Dimensions: 181 x 125 x 76 mm (7.1 x 4.9 x 3”)
4-20 mA ALARM LEVELS
Selectable 2 mA & 3.5 mA options
NB: All specifications are against an isobutylene calibration at 20 °C, 90% RH and up to 100 ppm.
* The Fence Electrode technology referred to in this document is produced by Ion Science Ltd, and
protected by patents. U.S. Patent No. 7,046,012. EP 1474681, other patents pending. ** TVOC has a
variable duty cycle set to 1 minute as standard. The user can adjust this to between 5 seconds
and 5 minutes.
Accessories
TVOC is available with an exclusive
range of accessories. Contact Ion
Science for more information.
Dimensions
ION SCIENCE is ISO9001:2008 certified
For further information and ordering please call:
Tel: +44 (0) 1763 208 503
Email: info@ionscience.com
www.ionscience.com