HYDROSTEEL 6000
TECHNICAL SPECIFICATION
MEASUREMENT PRINCIPLE
Hydrogen collection and analysis
INTRINSICALLY SAFE APPROVALS
II 2 G EEx ia IIC T4 -20 °C Ta 60 °C
Baseefa 02ATEX0093
OPERATION
Battery Type: 4 x Alkaline AA. Typically 30 hours life DC input 6 volts
SENSITIVITY
± 1 pL/cm2/s (Note: 1 pL = 10-12 litres at s.t.p.)
DYNAMIC RANGE
0 to 15,000 pL/cm2/s
ACCURACY
± 10% display reading ± one digit
RESPONSE
T90 = 50 seconds
CALIBRATION
Certified to international UKAS/NIST standards
Yearly calibration required
COMMUNICATION
Bi-directional using IR port
DATA LOGGING
Over 40,000 data points with time & date stamp
PC SOFTWARE
Microsoft Windows and Vista compatible
TEMPERATURE
Analyzer: -20 to 50 °C, -13 to 120 °F
LT-R probe: steel surface up to 150 °C, 300 °F
HT-R probe: steel surface up to 500 °C, 930 °F
AT-S probe: steel surface up to 500 °C, 930 °F
HT-S probe: steel surface up to 500 °C, 930 °F
Storage < 20 °C, <70 °F
GAS CONDUIT
Standard lengths available:
LT-R and HT-R 1 - 10m
LT-R only 2m
DIMENSIONS
Analyser: 13.5 x 2.3 x 2 in., (34 x 6 x 5 cm)
LT-R and AT-S probes: 6 in., (15 cm) diameter
HT-R and HT-S probes: 2.25 in., (5.7 cm) diameter
WEIGHT
Analyser: 1 lb, 450 g
LT-R probe: 14 oz, 397 g
HT-R probe: 11 oz 312 g
STEEL CUVATURE RANGE
Minimum diameters
LT-R and HT-S probes: 3.5 in., 9 cm
HT-R and AT-S probes: 8 in., 20 cm
EMC tested EN50081-1 & EN50082-1 July 98
ION SCIENCE is ISO9001:2000 certified
(December 2003)
Part number HS6V1.1 This publication is not intended to form the basis of a contract and specifications can change without notice.
For further information and ordering please call:
Tel: +44 (0) 1763 208 503
Email: info@ionscience.com
www.ionscience.com
ION SCIENCE LIMITED
The Way, Fowlmere, Cambs. SG8 7UJ, UK
Hydrosteel technology is also available in a
fixed monitor, Hydrosteel 7000, used for
continuous corrosion measurements at
highly critical sites.
Easy to install and non-intrusive, Hydrosteel 7000 can be
permanently attached to steel in petrochemical service up to
500 °C, 930 OF. Data is stored internally and communicated via a
4-20 mA output every 10 minutes, ensuring immediate awareness
of any change in active corrosion.
Dependent upon application Hydrosteel 7000 can give a direct
corrosion rate indication in mm/yr or mils/yr. Specialised
software collates information regarding the hydrogen flux and
steel type, temperature and thickness to calculate a
corrosion rate in real-time.
|
|