SKY® Capacitance Diaphragm Gauge Vacuum Gauges
Vacuum Control Catalog 2009 - A13
Specifications (Profibus DP)
Specifications (DeviceNet™)
Specifications (mbar based products)
CDG100D (continued)
CDG100D Profibus DP
Baud rates kBaud 9.6 / 19.2 / 93.75 / 187.5 / 500
MBaud
1.5 / 12
Address address 00 - 125 by switch or network programmable
Digital functions read pressure, status, ID
set
set points, filter, zero adjust, factory reset, DC offset
Connector for Profibus DP D-sub, 9 pin, female
Connector for CDG (analog output, supply voltage, setpoints) D-sub, 15 pin, male
CDG100D DeviceNet™
Protocol DeviceNet™, group 2 slave only
Data rate kBaud 125, 250, 500 by switch or network programmable
Cable length
125 kbps m (ft) 500 (1650)
250 kbps m (ft) 250 (825)
500 kbps m (ft) 100 (330)
MAC ID address 00 - 63 by switch or network programmable
Digital functions read pressure, status, ID
set
set points, filter, zero adjust, factory reset, DC offset
Specification DeviceNet™ ”Vacuum Gauge Device Profile” (ODVA)
Device type “VG” vacuum gauge
I / O slave messaging polling only
Supply voltage for gauge at D-sub connector +14 … +30 VDC or ±15 V / 12 W
Supply voltage for DeviceNet transceiver at
microstyle connector 24 V nom / <2 W (11 … 25 V)
Connector for DeviceNet™ microstyle, 5 pin
Connector for CDG (analog output, supply voltage CDG, setpoints) D-sub, 15 pin, male
Measurement Range mbar 1100 100 10 1 0.1
F.S. (Full Scale) Pa 110,000 10,000 1,000 100 10
Accuracy1) % of reading 0.2 0.4
Temperature effect
on zero % F.S./°C 0.0025 0.005
on span % of reading/°C 0.02
Pressure, max. kPa (absolute) 400 260 130
Resolution % F.S. 0.003
Further specifications see table «SPECIFICATIONS (Torr based standard products)».
|