Vacuum Gauges SKY® Capacitance Diaphragm Gauge
Vacuum Control Catalog 2009 - A12
Specifications (Torr based standard products)
CDG100D (continued)
Measurement Range Torr 1000 100 10 1 0.1
F.S. (Full Scale) Pa 133,322 13,332 1,333 133 13
mbar 1333 133 13.3 1.3 0.13
Accuracy1) % of reading 0.2 0.4
Temperature effect
on zero % F.S./°C 0.0025 0.005
on span % of reading/°C 0.02
Pressure, max. kPa (absolute) 400 260 130
Resolution % F.S. 0.003
Lowest reading % F.S. 0.01
Lowest suggested reading % F.S. 0.05
Lowest suggested control
pressure % F.S. 0.5
Temperature
Operation (ambient) °C +10 ... +50
Bakeout at flange °C 110
Storage °C –40 ... +65
Supply voltage +14 ... +30 VDC or ±15 V (±5%)
Power consumption
During Heat up W 15
At operating temperature W 10
Output signal (analog) VDC 0 ... +10
Response time ms 30 130
Degree of protection IP 40
Standards EN 61000-6-2, EN 61010, UL 61010-1, CSA 22.2 No.61010-1, SEMI S-2
Electrical connection D-sub, 15 pole, male
Setpoint two setpoints (SP1, SP2)
Relay contact VDC / ADC 30 / 0.5
Hysteresis % F.S. 1
Diagnostic port
Protocol RS232-C
Read pressure, status, ID,
Set set points, filter, zero adjust, factory reset, DC offset
Materials exposed to vacuum Aluminum oxide ceramic (AI2O3), stainless steel (AISI 316L2)),
Nickel, sealing glass
1) Non-linearity, hysteresis, repeatability at 25°C ambient operating temperature without temperature effects after 2 hours operation.
2) 18% Cr, 10% Ni, 3% Mo, 69% Fe
Measurement Range Torr 500 200 50 20 5 2 0.5 0.25
F.S. (Full Scale) Pa 66,661 26,664 6,666.1 2,666 666.61 266.66 66.66 33.3
mbar 666.61 267 66.67 26.7 6.6661 2.67 0.67 0.33
Accuracy1) % of reading 0.2 0.4
Temperature effect
on zero % F.S./°C 0.0025 0.005
on span % of reading/°C 0.02
Pressure, max. kPa (absolute) 400 260 130
Resolution % F.S. 0.003
Further specifications see table above.
Specifications (Torr based other ranges)
|