1•16
GasBadge Plus
Personal Single Gas Monitor
• Low-cost CO, H
2 S, O
2 , NO
2 or SO
2 monitoring• 2-year continuous monitoring• PPM or % by volume readout• Extremely water resistant – third-party certified IP66/67• User adjustable alarm and calibration gas setpoints Monitor shown actual size SPECIFICATIONS
CASE:
Rugged, water-resistant polycarbonate shell with protective concussion-proof overmold. RFI resistant. >
IP RATING:
G
asBadge
® Plus provides low-cost, personal protec-tion from dangerous levels of carbon monoxide, hydrogen sulfide, oxygen, nitrogen dioxide or sulfur dioxide with a number of high-end features not found in other similarly priced instruments.At 2.5 ounces (72 grams), the two-year, maintenance-free monitor’s compact size and light weight allow it to be worn comfortably with a variety of clip attachments. With the top-mounted sensor, the unit can also provide continuous and unobstructed protection even when placed in a shirt pocket.The GasBadge
Third-party certified IP66/67 (water resistant). >
DIMENSIONS:
3.2" x 1.9" x 1.1" (81.3 mm x 48.3 mm x 27.9 mm) >
WEIGHT:
2.5 oz (72 g) >
SENSORS:
Oxygen and Toxic Gases – Electrochemical >
MEASURING RANGES:
Carbon Monoxide – 0 to 1,500 ppm in 1 ppm increments
Hydrogen Sulfide – 0 to 500 ppm in 0.1 ppm increments
Oxygen – 0 to 30% by volume in 0.1% increments
Nitrogen Dioxide – 0 to 150 ppm in 0.1 ppm increments
Sulfur Dioxide – 0 to 150 ppm in 0.1 ppm increments >
® Plus is housed in a rugged enclosure that is extremely durable and resistant to water and radio frequency interference. A concussion-proof overmold protects the unit from extreme abuse in a variety of harsh industrial environments. The large, LCD display includes a backlight for clear visibility in low-light conditions and international graphic symbols for easy operation. The GasBadge
Non-replaceable Lithium Battery
Maintenance-free operation for 2 years >
TEMPERATURE RANGE:
-40°F to 140°F (-40°C to 60°C) typical >
HUMIDITY RANGE:
® Plus can be configured to show both gas type and direct gas read-
ings, or just the gas type. The instrument’s two-button interface provides intuitive navigation and setup, which can be password-protected for added security. With user adjustable alarm and calibration gas set-points, the monitor can be customized to any industry or application. When gas concentrations exceed the preset limits, GasBadge
0 to 99% RH (non-condensing) typical >
APPROVALS
UL, cUL and CSA – Class I, Div 1, Groups A,B,C,D; T4
MSHA – Intrinsically safe for Methane/Air mixtures only
ATEX – EEx ia IIC T4
Equipment group and Category: II 1G
EMC: EN50270
Performance – O (EN50104); CO, H S, (EN45544) (pending)Australia – Ex ia I/IIC T4
IEC – Ex ia I/IIC T4
Russia – GOST-R Approval (pending) >
® 2 2 Plus alerts the user with a powerful audible alarm complemented by vibrating and visual alarms. The standard confidence indicator, automatic self
Continued >