You may also be interested in
Gas analyzer, Analyzer, Calibrator, Pressure calibrator
Text version of the page
300.165.000 | 10.07 Data instantly available on theFlowAnalyser Easy and simple Plug & Play
installation...
TM display ...through the RS-232 port. The imtmedical MultiGasAnalyserTMOR-703 probe is designed using the latest advances in microsystem technology. The MultiGasAnalyserTMOR-703 provides a complete in-line, real-time monito-ring system with unique versatility and design. When testing the performance and accuracy of anesthesia delivery and monitoring systems or CO2monitoring devices, the sensor proves its abso-lute reliability. The MultiGasAnalyserTMOR-703 sensor head measures infrared light absorption atseveral different wavelengths and determines the precise gas concentrations of the mixtures.Measurements CO2 Range0 - 10%Accuracy+/- 8% of reading or +/- 0.3%N 2 ORange0 - 100%Accuracy+/- 8% of reading or +/- 2%HAL, ISO, ENFRange0 - 5%Accuracy +/- 8% of reading or +/- 0.2%SEVRange0 - 8%Accuracy+/- 8% of reading or +/- 0.2%DESRange0 - 18%Accuracy+/- 8% of reading or +/- 0.2% Response Time CO 2 2 O and the 5 Anaesthetic Agents TM Numerical data and real-time curves available through the FlowLab TM softwareInterfacethrough RS-232 port Physical Data Weight MultiGasAnalyser OR-703 Brochure EN - 58469 OKPlease contact us directly for further information. We would be glad to assist you.imtmedical ag Gewerbestrasse 8 9470 Buchs SG Switzerland T: +41 81 750 66 99 F: +41 81 750 66 95www.imtmedical.com |
|