FlowAnalyser

TM

Technical Data

PF-300 PF-301 PF-302 Flow & Pressure Measurements
RangeAccuracy 300.718.000 | 08.08 FlowMeasuring directionbidirectional ••• Temperature compensatedyes ••• Pressure compensatedyes ••• Humidity compensatedyes ••• O
2 compensatedyes ••• High± 300 L/min± 1.75 %* or ± 0.1 L/min** ••• Low± 20 L/min± 1.75 %* or ± 0.04 L/min** ••• PressureHigh 0 - 10 bar± 1 %* or ± 10 mbar** ••• Average± 150 mbar± 0.75 %* or ± 0.1 mbar**
DifferenceRelativeRelative Low 0 - 5 mbar ± 1 %* or ± 0.01 mbar** • High in Flow Canal0 - 150 mbar ± 0.75 %* or ± 0.1 mbar** ••• Barometer0 - 1150 mbar (abs) ± 1 %* or ± 5 mbar** ••• Vacuum pressure± 1000 mbar ± 0.5 %* or ± 2 mbar** • Measuring unitFlowL/min, L/s, cfm, mL/min, mL/s ••• Pressurebar, mbar, cmH
2 O, inH
2 O, Torr, inHg, hPa, kPa, mmHg, PSI ••• Additional Measuring Values
RangeAccuracy OxygenConcentration0 - 100 %± 1 % O
2 ** ••• Pressure compensatedyes ••• TemperatureHigh in Flow Canal0 - 50°C± 1.75 %* or ± 0.5°C** ••• Dew pointHigh in Flow Canal-10 - 50°C± 2 %* or ± 1°C** ••• Air humidityHigh in Flow Canal0 - 100 %± 3 %** ••• CO
2 Concentration0 - 20 %± 8 %* or ± 0.3 %**
with OR-703with OR-703with OR-703 N
2 OConcentration0 - 100 %± 8 %* or ± 2 %**
with OR-703 with OR-703 with OR-703 HAL, ISO, ENFConcentration0 - 12 %± 8 %* or ± 0.2 %**
with OR-703 with OR-703 with OR-703 SEVConcentration0 - 15 %± 8 %* or ± 0.2 %**
with OR-703 with OR-703 with OR-703 DESConcentration0 - 22 %± 8 %* or ± 0.2 %**
with OR-703 with OR-703 with OR-703 Gas typesAir, Air/O
2 , N
2 O/O
2 , Heliox (21% O
2 ), He/O •••
2 , N
2 , CO
2 , customized gas types Gas StandardisationATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPD,0/1013, 20/981, 15/1013, 25/991, 20/1013 ••• Respiratory Parameters
1) RangeAccuracy Rate1 - 1000 bpm/min±1 bpm or ± 2.5 %** ••• TimeT
I ,T
E 0.05 - 60 s± 0.02 s ••• I:E ratio1:300 - 300:1± 2.5 %* ••• Ti/Ttotal0 - 100 %± 5%* ••• Breath volumesVti, Vte± 10 L± 2 %* or ± 20 mL** ••• Minute volumesVi, Ve0 - 300 L/min± 2.5 %* ••• PressureP
peak , P
mean , PEEP, P
plateau 0 -150 mbar± 0.75 %* or ± 0.1 mbar** ••• PeakflowPeakflow Insp./Exp.± 300 L/min± 1.75 %* or ± 0.1 L/min** ••• ComplianceC
stat 0 - 1000 mL/mbar± 3 %* or ± 1 mL/mbar** ••• TriggerAdult, Pediatric, HFOAdjustable on flow or pressure curves with user-defined limits. ••• General Information Electrical & AC input90 - 260 VAC, 50/60 Hz Physical DataBattery (lead rechargeable battery)3 hrs (with OR-703 2 hrs) ••• ••• Power consumption23 VA ••• Weight3.7 kg ••• Dimensions (w x d x h)22 x 25 x 12 cm ••• Data Storage a ll parameters (measured as well as respiratory values) ••• DisplayGraphic displayIntuitive user interface with numerical measuring values, statistics, volume trigger configuration, ••• gas type selection and calibration menus.CommunicationUSB for Windows Software FlowLab
TM , InterfacesRS-232 for individual communication, ••• TTL for external trigger.Calibrationannually ••• ConditionsAmbient temperature10 - 40°C (50 - 104°F) ••• Humidity10 - 95 % R.H.*** ••• ApprovalsCE, CSA •••
Legend *Tolerance related to the measured value ** Absolute tolerance