S02 analyzer adopts ultraviolet fluorescence method and combines with advanced microprocessor technology. It provides accurate and reliable detection of S02 at ppb~ppm level which meets the requirements of ambient gas monitoring and online monitoring of pollution source with dilution method. The operation principle is based on ultraviolet fluorescence method. After absorbing the ultraviolet energyg, S02 molecule is inspired to excited state and once it returns to ground state, it will release specific fluorescence. The intensity of fluorescence could realize quantitative measurement. Measurement stability is greatly improved by using timed dark current measurement to compensate photomultiplier drift and reference detector to calibrate the change of UV light intensity. With high-efficiency hydrocarbon remover and unique optical design the analyzer can effectively clear up inexactitude caused by interference. This analyzer can completely control all functions and also provides online instruction for important work parameters. Real time auto compensation for temperature and pressure will be made during measurement. Features 0 Measuring range: 0-50ppb to 0-20ppm,(selectable). 0 Dual measurement range and auto set range. 0 Multiple functional operations by microprocessor. 0 Task software is allowed to monitor test data during operation. 0 Continuous self-checking with alarm function. 0 Self-check of light intensity attenuation. 0 Bidrection RS-232 port for remote control. 0 Digital output working parameter. 0 Response time optimized by signal filter technology of self-adaptation. 0 Compensation function for temperature and pressure. 0 USB and Ethernet interface. 0 Colorful display and screen touch operation. 0 Large memory and history data saved automatically. Specifications Parameter Unit Measurement range Zero noise Lower detection limit Span noise Indicator error 20% F.S. accuracy 80% F.S. accuracy 24h Zero drift 24h 20%Span drift 24h 80%Span drift Response time Long-term zero drift Long-term span drift MTBF Work condition Dimension & weight Power supply Ports Protocol £ET/AN Process & Environmental Hangzhou Zetian Technology CO., Ltd E-mail: export@zetian-tech.com TEL: 0571-28322780 Add: Ind Building Taiheng Mansion,N0.22 Zhiren Street,Binjiang District,Hangzhou, Zhejiang,China (310053)

Open the catalog to page 1
zetian/environmental/air quality monitoring system/gas analyzer/analyzer - 2

NOx analyzer adopts chemiluminescence method and combines with advanced microprocessor technology. It provides accurate and reliable detection of NOx at ppb~ppm level which meets the requirements of ambient gas monitoring and online monitoring of pollution source with dilution method. The operation principle is based on chemiluminescence method. After absorbing the chemical energy and the reacting with 03 NO is inspired to excited state and once it returns to ground state, it will release energy in form of light quantum. The intensity of light emission is directly proportional to gas...

Open the catalog to page 2
zetian/environmental/air quality monitoring system/gas analyzer/analyzer - 3

CO analyzer adopts non-dispersive infrared spectrum and combines with advanced microprocessor technology. It provides accurate and reliable detection of CO at ppm level to meet the requirement of ambient gas monitoring. The principle is non-dispersive infrared based on gas filter correlation technology. CO molecule absorbs specific infrared light which is consistent with Beer-Lambert principle. CO concentration can be obtained by comparing infrared energy adsorption by sampling gas and reference gas. This analyzer can completely control all functions and also provides online instruction for...

Open the catalog to page 3
zetian/environmental/air quality monitoring system/gas analyzer/analyzer - 4

Ozone analyzer adopts ultraviolet spectrometry principle and combines with advanced microprocessor technology. It provides accurate and reliable detection of 03 at ppb~ppm level to meet the requirement of ambient gas monitoring. UV adsorption Ozone analyzer is developed based on Beer-Lambert technology. It is applied to monitor low concentration 03 in ambient air. 254nm UV light passes through sample gas cell and is absorbed by ozone inside it. Adsorption value has a certain ration with 03 concentration. By cyclically alternating of magnetic valve and measuring sample gas and the sample...

Open the catalog to page 4
zetian/environmental/air quality monitoring system/gas analyzer/analyzer - 5

Based on microprocessor technology, Dynamic Dilution Calibrator is developed for precise gas analyzer and supports multi paths calibration. With high-accuracy mass flow controller and standard gas, this system can provide calibration gas for zero point and span inspection of multiple gas components. Up to 4 gas sources can be provided. Built-in programmable ozone generator is optionally equipped with dynamic dilution calibrator. It can be used to calibrate 03 accurately and reliably. Also It can be used in GPT to produce N02. Linear drive of multi-point section is adopted for ozone...

Open the catalog to page 5
zetian/environmental/air quality monitoring system/gas analyzer/analyzer - 6

The zero gas generator provides clean and dry zero-level air. It is mainly applied to zero calibration of ambient air quality monitoring system or analyzer. It can also be used as flushing gas in permeation tube & furnace and oxidant gas for FID analyzer. Zero gas generator includes external oil-free air compressor, pressure and flow controller, water removal system. Based on adsorption, filter, reaction and other principles, the removers can remove S02, NO, N02, 03, H2S, CO, NH3, hydrocarbon, particulate matter and output dry and clean air as well. The water removal system inside the...