MCP ASSEMBLY WIRING EXAMPLES MCP ASSEMBLY WIRING EXAMPLES
Any of the MCP-IN, MCP-OUT, anode and phosphor screen can be at ground potential.
Signal Detection
● Positive ion detection (Used with high-voltage power supply only) ● Electron or negative ion detection (Used with high-voltage power supply and divider circuit)
MCP (2-STAGE) MCP (2-STAGE) Amp.C(3 kV)
CR0.1 M >
A
2 M Ω Ω -0.1 kV-2 kV MAX.
ANODE +2.1 kV MAX.
ANODE >
TMCPC0005EF ● Positive ion detection (Used with high-voltage power supply only) ● Electron or negative ion detection (Used with high-voltage power supply and divider circuit)
MCP (2-STAGE) MCP (2-STAGE) PHOSPHOR SCREENPHOSPHOR SCREEN 2 M Ω 4 M Ω +6 kV Max.-2 kV Max.+4 kV Max. >
TMCPC0007ED 13