You may also be interested in
Hydrogen detector, Hydrogen analyzer, Leak detector, Analyzer, Detector
Text version of the page
HY-OPTIMA™ 730 Process Hydrogen Analyzer
• CO and H2S Tolerance
Description
H2scan’s Series 700 in-line, real-time hydrogen specific process analyzers are designed for ease of use, interface
flexibility and true process control. The HY-OPTIMA™ 730 is a solid-state sensor configured to operate in background
gas streams having carbon monoxide concentrations up to 100 ppm, sulfur stream concentrations up to 100 ppm, up to
95% relative humidity, and temperatures up to 100ºC. The HY-OPTIMA™ 730 is ideal for hydrogen production and
petrochemical applications where real-time measurements of hydrogen can increase process plant efficiencies,
improve diagnostics and enhance maintenance management translating into higher profitability.
Operating Conditions
Process Gas Tolerance:
CO tolerance < 100 ppm
H2S tolerance < 100 ppm
Operating Humidity: 0% to 95% RH
Flow Rate: 0.1 slpm to 50 slpm
Pressure: 0 to 7 bar gauge, 0 to 100 psi gauge
Process Gas Temperature Range: -20ºC to 100ºC
Operating Temperature Range: -20ºC to 40ºC
Storage Temperature Range: -40ºC to 50ºC
Calibration Background Gas: Nitrogen
Adapter Fittings
½ in. MNPT thread
½ in. FNPT thread
-8 SAE/MS thread size
... and many other industry standards
Dimensions
Total Length: 9.1 in.
Length of Enclosure: 6.7 in.
Length of Tube: 2.4 in.
Width: 3.4 in.
Depth: 1.4 in.
Certifications:
.
Performance
Hydrogen Sensitivity Range:
0.5% to 100% hydrogen at 1 ATM on a
volume basis in a single or multi-component
gas background
Typical Response Time: T90 less than 30 sec
Ingress Protection: IP64 capable
Recommended Verification Interval: 90 days
Product Life Expectancy: 10 years
Accuracy(*):
± 0.15% absolute for 0.5 to 10% H2
± 0.50% absolute for 10 to 100% H2
Drift/week:
± 0.10% absolute for 0.5 to 10% H2
± 0.20% absolute for 10 to 100% H2
Repeatability:
± 0.10% absolute for 0.5 to 10% H2
± 0.20% absolute for 10 to 100% H2
Linearity:
± 0.10% absolute for 0.5-10% H2
± 0.20% absolute for 10-100% H2
Interface Options
Input Voltage Range: 8 VDC to 13 VDC
Input Current: 1.0 A
Analog Outputs:
Output Voltages - OR - Output Currents
• 0 VDC to 5 VDC • 4 mA to 20 mA
• 1 VDC to 5 VDC • 0 mA to 20 mA
• 0 VDC to 4 VDC • User-specific mA range
• 0.5 VDC to 4.5 VDC
• User-specific VDC range
Serial Communication Options: RS232 or RS422
Relay Contacts: Two programmable 60 VDC / 1 A
SPDT relays with both normally open (N.O.) and
normally closed (N.C.) contacts
HY-OPTIMA™ is a registered Trademark of H2scan.
Specifications subject to change without prior notice.
H2scan Sept. 2009, International Release
90000017 R4
28486 Westinghouse Place, Ste.100, Valencia, CA 91355
Tel: 661-775-9575, Fax: 661-775-9515
Email: sales@h2scan.com
Website: www.h2scan.com
Sensor performance specifications are only valid for units configured for a maximum 65ºC dry process stream
temperature. All figures assume pressure compensation, operating in ambient that do not contain Oxygen and
are in addition to any errors in the gasses used. The accuracy is specified for serial port output only.
|
|