HY-OPTIMA™ 1730 Process Hydrogen Analyzer - H2scan - #1

/ 2


catalogue search
HY-OPTIMA™ 1730 Process Hydrogen Analyzer - H2scan
P. 01
HY-OPTIMA™ 1730 Process Hydrogen Analyzer - H2scan
P. 02
Pages:
HY-OPTIMA™ 1730 Process Hydrogen Analyzer - H2scan


See other catalogues for H2scan

Text version of the page
HY-OPTIMA™ 1730 Process Hydrogen Analyzer Description H2scan’s HY-OPTIMATM 1730 in-line, real time hydrogen-specific process analyzers are designed for ease of use, interface flexibility and true process control. The HY-OPTIMA™ 1730 is an intrinsically safe device that incorporates our patented solid-state sensor and is configured to operate in background gas streams having carbon monoxide concentrations up to 100 ppm, sulfur stream concentrations up to 100 ppm, up to 95% relative humidity and temperatures up to 100ºC. The HY-OPTIMA™ 1730 is ideal for hydrogen gas production, petroleum refining and petrochemical applications where continuous real time measurements of hydrogen can increase process plant efficiencies, improve diagnostics and enhance maintenance management translating into higher profitability. Performance Hydrogen Sensitivity Range: 0.5% to 100% hydrogen at 1 ATM Response Time: T90 less than 30 sec Ingress Protection: IP64 capable Recommended Verification Interval: 90 days Product Life Expectancy: 10 years Accuracy(*): ± 0.15% absolute for 0.5 to 10% H2 ± 0.50% absolute for 10 to 100% H2 Drift/week: ± 0.10% absolute for 0.5 to 10% H2 ± 0.20% absolute for 10 to 100% H2 Repeatability: ± 0.10% absolute for 0.5 to 10% H2 ± 0.20% absolute for 10 to 100% H2 Linearity: ± 0.10% absolute for 0.5-10% H2 ± 0.20% absolute for 10-100% H2 Interface Options Power Barrier Input Voltage: 24 VDC nominal Power Barrier Input Voltage Range: 20 VDC to 28 VDC Intrinsically Safe Sensor Input Voltage: 10 VDC Sensor Input Voltage Range: 5 VDC to 28 VDC Input Power: 10 W Analog Output: 4 mA to 20 mA Serial Communication Options: RS422 Certifications 0359 II 1 G, Ex ia II H2, d+d IIB T3 II 2 G, Ex ib II H2, ib d IIB T3 ITS07ATEX25634X Operating Conditions Operating Humidity: 0% to 95% RH Flow Rate: 0.1 slpm to 50 slpm Pressure: 0 to 7 bar gauge, 0 to 100 psi gauge Process Gas Temperature Range: -20ºC to 100ºC Operating Temperature Range: -20ºC to 40ºC Storage Temperature Range: -40ºC to 50ºC Calibration Background Gas: Nitrogen Included Accessories Intrinsically Safe Power Barrier Armored Power / Analog Cable: 4 m Calibration Cup Optional Accessories Single Channel Intrinsically Safe Analog Output Barrier (with or without relays) Single Channel Intrinsically Safe RS422 Serial Interface Barrier Serial Interface Cable: 4 m (custom lengths available) Armored Power / Analog Cable (custom lengths available) Adapter Fittings ½ in. MNPT thread ½ in. FNPT thread -8 SAE/MS thread size ... and many other industry standards. Dimensions Total Length: 10.3 in. Length of Enclosure: 6.8 in. Length of Tube: 2.9 in. Width: 3.4 in. Depth: 1.4 in. Sensor performance specifications are only valid for units configured for a maximum 65ºC dry process stream temperature. All figures assume pressure compensation, operating in ambient that do not contain Oxygen and are in addition to any errors in the gasses used. The accuracy is specified for serial port output only. HY-OPTIMA™ is a registered Trademark of H2scan. Specifications subject to change without prior notice. H2scan September 2009, International Release 900000048 R0 28486 Westinghouse Place, Ste.100, Valencia, CA 91355 Tel: 661-775-9575, Fax: 661-775-9515 Email: sales@h2scan.com Website: www.h2scan.com

pageCatalog pdf di En 2012-05-20-16