HYDROGEN SPECIFIC LEAK DETECTION OPERATING MANUAL Hydrogen is the future, we can sense it! 27215 Turnberry Lane, Unit A

 Open the catalogue to page 1

HY-ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL Deliver unsurpassed value and optimize green initiatives with our one of a kind continuous hydrogen-specific sensing technology worldwide. Enable end-user customers to efficiently and effectively optimize:  Electric utility power transformer fleet and other oil-filled assets (Grid)  Petroleum refinery and other industrial process control  Facility and equipment safety to minimize downtime ...at a much lower total costs of ownership than the competition. H2scan's technology accepted as the new gold standard in hydrogen sensors....

 Open the catalogue to page 2

HY-ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL

 Open the catalogue to page 3

HY-ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL IMPORTANT NOTICES Read and understand this operating manual before installing or using the unit. Only use cables, battery pack, battery charger, and AC/DC power supply from H2scan with this unit. If this equipment is used in a manner not specified by H2scan, the protection provided by this equipment may be impaired. Hydrogen is flammable at 4% in air. Take indications seriously and be prepared to take action. In the event of detection of 4% or higher of a hydrogen gas concentration there is a high probability of a hazard to safety....

 Open the catalogue to page 4

HY-ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL 1 DESCRIPTION H2scan believes that protecting lives means being able to locate and find hydrogen leak as quickly as possible. With two sensing elements on the same semiconductor die, the HY-ALERTA™ 500 can detect hydrogen leaks as low as 15 ppm and will not saturate or be destroyed when detecting concentrations of hydrogen up to 100%. The flexible cable allows the sensor probe access to virtually all potential leak sources. SPECIFICATIONS Sensitivity Range: Response Time: Indication of hydrogen within seconds. Stabilization to final...

 Open the catalogue to page 5

HY-ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL Sensor Probe Controller Coiled Cord Controller Carry Pouch 90000002 R12 Battery Charger with Regional Plugs Page 6 of 14

 Open the catalogue to page 6

HY-ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL OPERATION 3.1 STARTUP To power-up the HY-ALERTA™ 500, press and hold the ON/OFF button until the Controller LCD display indicates an operational message. Warning: Only power-up the instrument in a hydrogen-free environment. After power is on, the instrument takes about ten minutes to warm-up. During this time the LCD displays a countdown to completion and the Probe Tip LED is amber. The following operations occur: The Wide Range Sensor® reaches operating temperature. A system self-test is run. Upon successful completion of the above...

 Open the catalogue to page 7

HY-ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL The Probe Tip LED Indicator shows an increase or decrease in hydrogen concentration. Leak detection is accomplished by watching the Probe Tip LED and the bar graph array and moving the sensor around a potential hydrogen leak. Probe Tip Colors Solid Green Pulsing Green - Unit ready - < 15 ppm hydrogen - Decreasing H2 level Solid Amber - System startup - Steady H2 level Pulsing Red Alternating Red, Green - Increasing H2 level - Fluctuating H2 level The number of yellow LEDs lit in the Probe LED bar graph array indicates detected hydrogen...

 Open the catalogue to page 8

HY-ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL 4.2 HYDROGEN-FREE AREAS For the purposes of this document a hydrogen-free area is one with less than 5ppm of hydrogen in the air. It may be difficult to find a hydrogen-free area in a facility where hydrogen is used. Nearby rooms, or even the entire building, may not be hydrogen free. To check these areas reset or zero the sensor outside, away from any hydrogen tanks, pipes, or other potential sources. Walk through the facility, watching the sensor. It is surprising how far low levels of hydrogen can spread. If the sensor is zeroed...

 Open the catalogue to page 9

HY-ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL Key ENTER ▲ ▼ ► ◄ ◄► Function Go to the Information Display menu. Display the peak hydrogen reading. No change Clear the peak hydrogen value. Zero the sensor. Reset the sensor. 5.3 GENERAL KEYPAD FUNCTIONS Navigation Enter submenu Previous Menu Next Menu Enter Submenu Exit Submenu (Back) Exit Configuration 5.4 INFORMATION DISPLAY The Information Display menu allows the user to view useful information about the instrument including firmware revisions, serial number, and calibration date. Enter it by pressing and holding the ENTER...

 Open the catalogue to page 10

HY-ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL The hydrogen plume from a leak generally spreads in a roughly conical shape that is easily disturbed by environmental conditions. Certain conditions such as pressure, temperature, and leak size may act together to change the shape of the hydrogen plume from a cone to a laser-like beam. This makes finding a leak more difficult. If the sensor element is near (and above) the leak, the concentration will likely be higher but the leak may be difficult to locate. As hydrogen dissipates the concentration decreases. Generally, greater distances...

 Open the catalogue to page 11

HY-ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL Symptom Possible Cause The probe is disconnected from Turn off the instrument and verify that the the controller. probe is properly connected to the controller. Error 88 Faulty capacitor Cycle of power The sensor PCB temperature is Error 40 Turn off the instrument. Let it cool. too high. The sensor temperature is Error 20 Cycle of power incorrect. Unit won’t turn on The battery is dead. Charge the battery completely. If the recommended action does not solve the problem, the HY-ALERTA 500 should be returned to the factory for repair....

 Open the catalogue to page 12

HY-ALERTATM 500 HANDHELD HYDROGEN LEAK DETECTOR OPERATING MANUAL Zero Menu Zero Sensor Continue? ENTER Reset Menu Reset Sensor Continue? Zero Sensor Not Done Zero Sensor In Progress Reset Sensor Done Reset Sensor Not Done Reset Sensor In Progress Zero Sensor Done

 Open the catalogue to page 13