You may also be interested in
Thermometer, Energy meter, Multimeter, Chart recorder, Handheld thermometer
Text version of the page
3-349-390-03
1/11.06
GMC-I Gossen-Metrawatt GmbH
METRATHERM.1
Infrared Thermometer with Laser Sight
for the Contactless Measurement of Surface Temperatures
Applications
• Temperature measurement at large distances, and thus at a
safe distance from potential danger zones
• Electrical troubleshooting
• Automotive repair and maintenance
• Air conditioning
• Scientific experiments
• Manufacturing processes in the field of semiconductor
technology
• Contacts in electrical switch boxes
• Food monitoring and processing
Applicable Regulations and Standards
Characteristic Values
IEC 61000-4-2 Electrostatic discharge immunity test
IEC 61000-4-3 Radiated radio-frequency electromagnetic field immunity
test
IEC 61000-4-8 Power frequency magnetic field immunity test
DIN EN 61326:2002 Electrical equipment for measurement, control and
laboratory use – EMC requirements
Quantity Characteristic Value
Infrared thermometer sensor
Temperature range –50 to 1000 °C (–58 to 1832 ºF)
Accuracy ±3.0 º C (±5 ºF) from –50 to –20 ºC (–58 to –4 ºF),
±2.0 ºC (±3 ºF) from –20 to 100 ºC (–4 to 212 ºF),
±2% from 100 to 1000 ºC (212 to 1832 ºF)
Wavelength 8 to 14 ìm
Repetition accuracy ±1 ºC or ±2 ºF
Resolution, main and auxiliary displays 0.1 ºC
Response time 500 ms
Emissivity adjustable from 0.1 to 1.0
Relationship of the distance between the
measuring instrument and the object to be
measured, and diameter of the spot (distanceto-
spot-ratio D:S)
50:1
Type K thermocouple sensor
Temperature range / accuracy –200 to +200 °C / ±(0.2% rdg. + 1.5 °C)
+200 to +1370 °C / ±(0.5% rdg. + 1 °C)
Temperature range / resolution –200 to –100 °C / 1 °C
–99.9 to +1370 °C / 0.1 °C
Supply power / battery service life 9 V block battery (6LR61) / 10 hours
Operating temperature 0 to 50 °C (32 to 122 °F),
10 to 90% atmospheric humidity
Selectable unit of measure: °C or °F .
Automatic shutdown Approx. 30 seconds after last key activation
Switchable laser beam .
Min. /max. and average values .
10 memory locations .
Acoustic warning signal .
USB data port .
Dimensions 200×127×47 mm
Weight Approx. 350 grams with battery
• Infrared temperature measuring range: –50 to 1000 °C (–58 to 1832 ºF)
• Exact positioning with laser sighting (can be activated if desired)
• High distance-to-spot (D:S) ratio: 50:1
• Emissivity adjustable from 0.1 to 1.00 in steps of 0.01
• Additional measuring option with type K thermocouple
• Illuminated LCD panel
• Selectable unit of measure: °C and °F
• Single and continuous measurements
• Storage of current measured temperature, lowest and highest
temperature or maximum temperature difference ÄT
• 10 memory locations for individual measurements
• Alarm function: Acoustic signal in the event of violation of
adjustable limit values
• USB interface for data transmission to a PC
• PC software for displaying and recording measured values
(data logger function) on included CD ROM
|
|