You may also be interested in
Ohmmeter, Insulation tester, Measuring device, Power meter, Infrared thermometer
Text version of the page
3-348-810-03
4/1.07
GMC-I Gossen-Metrawatt GmbH
METRAOHM.413
Low-Resistance Ohmmeter
Applications
The METRAOHM.413 is a battery powered instrument for the measurement
of low-value resistances in electrical systems.
Large protective conductor, earthing, equipotential bonding and
lightning protection systems can also be tested quickly and safely
with the METRAOHM.413. Resistance is measured between
reference earth (e.g. an equipotential bonding strip) and any
desired points to this end.
A measuring current of 200 mA and automatic functions assure
reliable measuring results.
– Adjustment of shunts in instrumentation applications
– Testing of electrical connections at busbars in open-pit
mining, in industry and in household applications
– Testing of cable resistance, wiring, shunt resistors in printed
circuit boards and thick-film circuits
– Measurement of contact resistance in relays, power
contactors and load interrupters
– Measurement of fuse resistance, as well as cable resistance in
heavy current circuits
– Testing of coil resistance in transformers, coils, small motors
etc.
Description
Easy Operation
Operation is very easy. Measurement cables of almost any length
can be used thanks to the offset balancing function.
Measurement polarity can be reversed, and both measured
values are stored to memory for subsequent comparison.
Resultant deviations in the measuring results (e.g. possible
galvanic voltages) can thus be recognized.
Offset Balancing
Offset balancing compensates for measurement cable resistance.
The offset value is saved to memory and is utilized for all future
measurements until balancing is performed once again. The value
remains in memory, even after the measuring instrument has been
switched off.
A variety of measurement cables with resistances of up to 3.5 Ů
can be used.
Protection Against Operator Error
The METRAOHM.413 is protected against inadvertent connection
to devices under test with fault voltages of up to 500 V by means
of protective devices.
Interference Voltage Warning
Interference voltage is indicated visually with a red LED, and
audibly with an acoustic warning signal. The voltage value is
displayed at the LCD as well, as long as the test probe makes
contact.
• Handy measuring instrument per VDE 0413 part 4 / EN 61557-1 / -4
• Large measuring range from 0.01 to 1999 kŮ
• Measurement cable zero balancing
• Compact and durable
For rugged service calls and laboratory use
• Overvoltage protection
Protects the instrument in the event of inadvertent connection
to mains voltage
• Indication of interference voltage
• IP 65 protection
|
|