Linear Range: 27.9 mm (1100 mils). Linear range begins at approximately 1.3 mm (50 mils) from target and is from 1.3 to 29.2 mm (50 to 1150 mils) (approximately –1.5 to –12.5 Vdc). Average Scale Factor (ASF) 394 mV/mm (10 mV/mil) nominal Deviation from best fit straight line (DSL) >
5 @ T >
a = -35 C to +85 C. Europe: Less than ±0.74 mm (±29 mils) System performance over extended temperatures: PENDING: EExia IIC T5 for Zones 0, 1 and 2, Group IIC, Baseefa (2001) Ltd. certificate number BAS99ATEX1101, when installed with intrinsically safe zener barriers or
galvanic isolators, T5 @ Ta= -35 Over a probe temperature range of -35°C to +120°C (-31°F to +248°F) with the Proximitor® Sensor and extension cable between 0°C to +45°C (+32°F to +113°F), the DSL remains within ±2.03 mm (±80 mils). Over a Proximitor® Sensor and extension cable temperature range of -35°C to +65°C (-31°F to +149°F) with the probe between 0°C to +45°C (+32°F to +113°F), the DSL remains within ±2.03 mm (±80 mils). C to +85 C. EEx nA for Zone 2, Group IIC, Baseefa (2001) Ltd. certificate number BAS99ATEX3100U. Recommended Minimum Target Size: 102 mm (4.0 in) diameter (flat target) Electrical Classification: >
Mechanical
Probe Tip Material: Complies with the European CE mark. Polyphenylene Sulfide (PPS). Probe Case Material: AISI 304 stainless steel (SST). Probe Cable Specifications: 75 triaxial, perfluoroalkoxyethylene (PFA) insulated FluidLoc probe cable in the following total probe lengths: 1, 5 or 9 metres. Extension Cable Material: 75 triaxial, perfluoroalkoxyethylene (PFA) insulated FluidLoc cable. Proximitor® Sensor Material: A380 aluminum Sliding Bracket Material: >
Hazardous Area Approvals
Anodized aluminum and stainless steel Multiple approvals for hazardous areas certified by Canadian Standards Association (CSA/NRTL/C) in North America and by Baseefa (2001) Ltd. in Europe are pending. North America: PENDING: Ex ia IIC T5; Class I Zone 0 or Exia IIC T5 for Class 1 Division 1; Groups A, B, C, and D, when installed with intrinsically safe zener barriers per drawing 141092 or when installed with galvanic isolators. ExnA IIC T5 Class I Zone 2 or ExnA IIC T5 for Class I, Division 2, Groups A, B, C, and D when installed without barriers per drawing 140979. T System Length: 5 or 9 metres including extension cable Probe and Extension Cable Armor (optional): Flexible SST with PFA outer jacket. >
For Reference Only – See Document #174014 Specifications and Ordering Information Page 3 of 21