You may also be interested in
Gas analyzer, Monitoring system, Hand-held analyzer, Analyzer
Text version of the page
Oxygen AnalyzerGeneral parametersGASMET ZrO2Oxygen AnalyzerGASMET 19” frame analyzer with microprocessor technology has the ZrO Measuring principle: ZrO2 measuring cell 2 -cell inside and can be used for O 2 -measurement in CEM-cabinets or as a portable analyzer. Together with a sampling and conditioning system GASMET Oxygen Analyzer can measure the O Detection limit: 2 Response time: 2 content in wet flue gas or steam down to less than 1 ppm O Environm. Temp.: -20 C to +40 C 2 . GASMET Oxygen Analyzer is manufactured from ENOTEC components. Line Voltage: 115 or 230 V / 50 -60 Hz Reference gas: instruments air, dew point less than –40 C, no oil, about 30 l/h Measuring gas: dry or wet, no combustibles Calibration gas: instrument air as above or test gas from bottles |
|