| | Measuring field: | 0,0001-25% O2 by volume. | |
| | Precision: | ±0,5% of the theoretical value or 0,1% O2 (which ever is the greater). | |
| | Stability: | within 1%. | |
| | Response time (T90): | zirconia cell < 1", the hole system < 15" | |
| | Storage temperature: | probe -40 °C/+80°C. | |
| | Working temperature: | probe 500 -r 1600°C for the temperature of the process. Room temp.< 150 °C for the head of the probe; industrial converter and barrier: -20 °C/ +50 °C. | |
| | Outputs: | analog 4-20 mA 500 ohm or 0 - 10 V 10 mA linear on one of the following fields: 0-1999 ppm; 0-5%; 0-10%; 0-25% O2. If the fault alarm is activated the analog output is forced to 2 mAor0 V. | |
| | Alarms | Potential free (SPST) contacts, O2 high and low, fault (fail safe) service and calibration in progress. For all contacts 250 V, 1 A max. | |
| | | time constant adjustable between 0 and 900 sec. | |
| | Power supply: | 115V and 230 V ±10%, 50-60 Hz, 50 V.A max. Safety according EN 61010 | |
| | Protection: | Ex d IIC 2 GD c T6 IP 65 | |
| | EMC | According E 50081 and EN 50082 | |
| | Connecting cable: | 4x1.5 shielded. Max length 100 mt. | |
| | Probe dimensions: | Immersion lengths = 650, 500, 310 mm; Diameter = 27 mm; With suitable extension tubes 1500 mm maximum immersion length can be reached. | |
| | | Inox compression fittings for 6x4 tubing or W" NPT-F | |
| | Options: | Hart & Mod-bus protocols | |
| | | | |