Models 3911 Extensometer for Indirect Tensile Testing of Asphalt - Epsilon Technology - #1

/ 1


catalogue search
Models 3911 Extensometer for Indirect Tensile Testing of Asphalt - Epsilon Technology
P. 01
Page:
Models 3911 Extensometer for Indirect Tensile Testing of Asphalt - Epsilon Technology


See other catalogues for Epsilon Technology
You may also be interested in

Calibrator, Load cell, Extensometer, Signal conditioner


Text version of the page
MODEL

Models 3911 Extensometer for Indirect Tensile Testing of Asphalt - 33099 3911

Extensometers for Indirect Tensile Testing of Asphalt Epsilon

Technology Corp

This model meets many of the needs for testingasphalt core samples,in 4 and 6 inch diameters. It meets test method requirements for strain measurement developed under the U.S.Federal Highways SHRP program. Features O

RDERING

I

NFORMATION •

Model 3911 indirect tensile extensometer for asphalt, for testing perAASHTO TP31 and ASTM D4123.

Model 3911 Available Versions: Consult factory for additional ranges.

See the electronics section of this catalog foravailable signal conditioners and strain meters.

Measuring Range •

Full bridge, 350 ohm strain gaged design for compatibility with nearly any test system.

DESIGNATION

-0003 +0.030” (0.76 mm) -0006 +0.060” (1.50 mm)

All standard units meet existing ASTM class B-1 and ISO 9513, class 0,5 requirements for accuracy. -0008 +0.080” (2.00 mm)

Includes high quality foam lined case.

Rugged, dual flexure design for improved performance. Model Number 3911-________

Easy mounting with built-in breakaway feature should specimen fracture.

Example: 3911-0006: +0.060 inches (1.50 mm)measuring range 3911 extensometer S PECIFICATIONS

For indirect tensile testing, such as for resilient modulus,these extensometers measure the lateral deformation of specimens. They are self-supporting on the sample and clip on in seconds. The traditional way this deformation has been measured was prone to errors caused by slight rocking of the sample as it is loaded. The self-supporting design of the Model 3911 eliminates this problem. This unit will handle the dynamic pulse requirements of resilient modulus testing.The speed of mounting makes test set-up much faster thanwith the older method. The pivot mounting used ensures thatthe specimen contact pads meet the often uneven asphaltsurface as well as possible. The measuring range of 0.050 inches allows a wide range of testing, yet can easily measure the sometimes very small displacements required.The Model 3911 extensometers are strain gaged devices,making them compatible with any electronics designed for strain gaged transducers. Most often they are connected to a test machine controller. The signal conditioning electronics for the extensometer is typically included with the test machine controller or may often be added. In this case the extensometer is shipped with the proper connector and wiring to plug directly into the electronics. For systems lacking the required electronics, Epsilon can provide a variety of solutions, allowing the extensometer output to be connected to data acquisition boards, chart recorders or other equipment.

Excitation: 5 to 10 VDC recommended, 12 VDC or VAC max. Output: 2 to 4 mV/V nominal, depending on the model Linearity: ≤ 0.20% of full scale measuring range, dependingon model Temperature Range: Standard is -40 °C to +100 °C (-40 °F to +210 °F) Cable: Integral, ultra-flexible cable, 8 ft (2.5 m) standard Specimen Size: Adjustable to fit both 4 inch and 6 inch diameter specimens (100 mm and 150 mm)
O PTIONS Connectors to interface to nearly any brand test equipment Shunt calibration module (see page 96)
8.07205154.26.074.37111.5614.3 Contact Epsilon for your special testing requirements. TYPICAL 3911 DIMENSIONS:[mm] inches

66 67