3641 High Temperature Un-Cooled Clip-On Gages Epsilon
Technology Corp
Designed for fracture mechanics tests in environmental
Ordering Information
Features
• 3603 signal conditioner included. Provides high level analog DC voltage output with very low noise (0.1 mV typical on 0 to 10 VDC calibrated signal). 10 ft (3 m) long output cable included.
• Shipped fully calibrated with electronics (traceable to NIST) with user specified voltage output.
• All standard units meet existing ASTM requirements for accuracy.
• Sharp grooves per ASTM E1820, E813, and E399 for improved stability when mounted.
• Includes high quality foam lined case.
• Rugged design provides dynamic testing capability to 20 Hz or faster depending on version.
Model 3641 Available Versions: Note: other gage lengths and measuring ranges available on special order.
chambers where the entire gage must be exposed to
Gage Length
U.S.A.
-0020
0.200"
-0025
0.250"
-0030
0.300"
-0047
0.475"
-0050
0.500"
METRIC
-003M
3.0 mm
-006M
6.0 mm
-010M
10.0 mm
-012M
12.0 mm
-0125M
12.5 mm
Measuring Range
the heat.These capacitive sensor based clip-on gages
DESIGNATION
-100T
+0.100"
-250T
+0.250"
-500T
+0.500"
-030M
+3.0 mm
-060M
+6.0 mm
-100M
+ 10.0 mm
-120M
+ 12.0 mm
-125M
+ 12.5 mm
(or COD gages) may be used up to 540 °C (I000 °F)
without any cooling.
3641 with 0.1 inch gage length and 0.02 inches measuring range.
These gages are designed for testing at elevated temper­atures, much higher than strain gage based clip-on designs. They can be used for several standardized fracture test methods, including ASTM E399 for fracture toughness or E1820 and E813 for JIC and R-curve determination. The Model 3641 uses a high temperature capacitive sensor and can operate without any cooling at the upper temperature limit of most environmental chambers used in materials testing.
High temperature nickel alloy springs are used to create the force to hold the gage in place on the knife edge contacts. Stiff arms with a flexure design at the rear eliminates potential problems of high temperature creep experienced with other designs.
The clip-on gages are supplied with the Model 3603 signal conditioning electronics. The gage and electronics are factory calibrated. The analog output voltage is typically calibrated for 0 to 10 VDC (other ranges are available on request). This can be used directly by data acquisition systems. It usually can be routed to a DC input channel on most test machine con­trollers as well.
Specification!
Model Number 3641-
Input: Includes power supply for your country (specify) Output: User specified, +/-5 VDC or +/- 10VDC typical Linearity: <0.10% of full scale measuring range, depending on model
Temperature Range: Ambient to 540 °C (ambient to 1000 °F)
Cable: Triaxial ceramic fiber insulated cable 3 ft (1 m) plus 10 ft (2.5 m) extension cable Operating Force: Exerts 2 to 3 lbs (9 to 14 N), depending on model Environment: Recommended for elevated temperature testing in air or some other gases
Options
Industrial connectors to interface to nearly any brand test equipment Bolt-on knife edges
Example: 3641-010M- 100M: 10.0 mm gage length, +10.0 mm measuring range
Signal
Conditioning
Electronics.
Model 3603 included with all Model 3641 extensometers.
[18.7] .74
H MEASURING RANGE
Contact Epsilon for your special testing requirements.
3641 MEDIUM MEASURING RANGE
DIMENSIONS: [mm] inches
5I
DirectIndustry's Virtual Technical Library: PDF Catalogue | Technical Documentation | Brochure | Manual | Industrial directory | Specifications | Characteristics
Search Go
page 1 p.1