EC9811
Ozone Photometer
The EC9811 Ozone Photometer meets U.S.
EPA requirements for an ozone analyser and
an ozone primary transfer standard. The
EC9811 is a compact analyser and primary
transfer standard that requires no external
pumps or scrubbers.
The EC9811 utilises the Beer-Lambert law and
UV absorption detection to calculate ozone
concentration. O3 measurements are
automatically corrected for gas temperature/
pressure changes and can be displayed in units
of ppm, ppb, ěg/m3 or mg/m3.
• Built in high efficiency sample pump features low noise operation and automatic flow
correction. The sample pump provides up to 5 LPM of scrubbed zero air for both the
photometer reference cell and the ozone generator.
• Photometric dual-channel technique provides reliable measurements.
• Single, small glass measurement cell, low sample flow rates and no mirrors means that cleaning
and maintenance requirements are minimised.
• Built-in secondary ozone scrubber allows independent verification of primary scrubber
performance ensuring maximum accuracy and reliability.
• Automatic five point span calibration of ozone generator.
• Stored data can be retrieved via RS232, USB interface or the optional Ethernet connection and
uploaded to a TCP/IP network.
• Ethernet option facilitates data download from an analyser connected to the internet via a
standard web interface. This feature also supports remote access to instrument parameters and
the status output screen .
• Inbuilt data logger uses Flash ROM to store up to 175 days of 5 minute averaged
data.