EE22 EE22
Humidity / Temperature Transmitterwith interchangeable probes
Unique for the EE22 series are the interchangeable sensing
probes with connector.
The calibration data is stored in the probes, which are therefore
interchangeable and probe replacement does not affect the performance of EE22.
The outstanding accuracy over the entire temperature range is
based on very precise calibration methods and on the latest
microprocessor technology. Well-proven E+E
humidity sensor
elements ensure excellent long-term stability. For high temperature applications (up to +80°C / >
+176°F )or in caseof limited space availability, the sensing probes can be connected
to EE22 housing with cables (2m, 5m or 10m / polycarbonatehousing
metal housing >
6.6ft, 16.4ft or 32.8ft )without any repercussions for the overall accuracy of the instrument.Voltage 0 - 1 / 5 / 10V or current 4 - 20mA(2 wire) outputs are available, of which the temperature output can be
scaled according to the application (see ordering guide).EE22 is suitable for direct wall mounting and for installation onrails according to DIN EN 50022. The optional display indicates the actual RH and Tvalues. In the pharmaceutical and biotechnology industry a Loop-Calibration of the RH and Toutputs, recommended
by the FDA(Food and Drug Administration), can easily be
performed utilizing separate RH and Tprobes (Type: EE22-xFTx2x).The RH and Toutputs can be adjusted with push buttons on the printed circuit board. >
RH probe Tprobe RR
As useful accessories reference probes (incl. test report) representingfixed humidity and temperaturevaluesare available. They shall be installed instead of the measuring probes to checkfunction and accuracy of the evaluation unit.One probe simulates high humidity and low temperature, the other low humidity and high temperature, to check the upper and lower end
of both analogue outputs. >
TTAA
pharmaceutical industry
clean rooms
storage rooms
green houses
cooling chambers interchangeable probesremote sensing probe up to 10m
(32.8ft) measuring range 0...100% RH / -40...80°C
(-40...176°F) optional displaytraceable calibrationcost saving, easy loop-calibration of RH and T probes
v1.4 EE22