TT D D EE0E EE0E
Measuring valuesRelative humidity
Humidity sensor >
1) 1) HC1000-400 Measuring range >
2) 0...100% RHAccuracy (including hysteresis, non-linearity and repeatability, traceable to international standards, administrated by NIST, PTB, BEV...)-15...40°C >
(5...104°F) <90% RH± (1.3 + 0.3%*mv) % RH-15...40°C >
(5...104°F) >90% RH± 2.3% RH-25...70°C >
(-13...158°F) ± (1.4 + 1%*mv) % RH -40...180°C >
(-40...356°F) ± (1.5 + 1.5%*mv) % RH Temperature dependence electronicstyp. 0.08% RH/°C Response time with filter at 20°C / t >
90 < 30 sec. >
Temperature
Temperature sensorPt1000 (DIN EN 60751, class A)
Measuring range sensor headEE30EX-A-20...60°C >
(-4...140°F) EE30EX-D -40...180°C >
(-40...356°F) EE30EX-E-40...180°C >
(-40...356°F) Accuracy (typ.) >
∆ °C °C Temperature dependencetypical 0.005°C/°C >
Max. adjustable Measurement Range
3) fromtounit EE30EX-AEE30EX-D/E HumidityRH0100100%RH TemperatureT-40 >
(-40) 60 >
(140) 180 >
(356) °C >
(°F) Dew point temperatureTd-80 >
(-112) 60 >
(140) 100 >
(212) °C >
(°F) Frost point temperatureTf-80 >
(-112) (32) (32) °C >
(°F) Wet
industrial bulb temperatureTw0 >
(32) 60 >
(140) 100 >
(212) °C >
(°F) Water vapour pressuree0 >
(0) 200 >
(3) 1100 >
(15) mbar >
(psi) Mixing ratior0 >
(0) 425 >
(2900) 999 >
(9999) g/kg >
(gr/lb) Absolute humiditydv0 >
(0) 150 >
(60) 700 >
(300) g/m >
3 (gr/ft 3 ) Specific enthalpyH-50 >
(-15000) 400 >
(150000) 2800 >
(999999) kJ/kg >
(lbf/lb) Outputs
Two freely selectable and scalable outputs0 - 5 V-1 mA< I >
L < 1 mA0 - 10 V -1 mA< I >
L < 1 mA 4 - 20 mAR >
L < 360 Ohm Serial interfaceRS232C >
General
Supply voltageSELV24V DC/V AC ± 15% >
SELV = Safety Extra Low Voltage Current consumption ≤ 150mA(24V DC); ≤ 280mA(24V AC) Pressure range with pressure tight sensor probe0.01...15 bar >
(0.15...218psi) System requirements for softwareMS Windows >
TM 98
or later; serial interfaceHousingssupply- and evaluation unitABS-plastic / IP65 sensor driver unit AlSi12 / IP65 Cable glandPG 7 and PG 9; for cable diameter 5 - 9 mm >
(0.2 - 0.35”) Electrical connectionscrew terminals max. 1.5 mm² >
(AWG 16) Sensor protectionsintered stainless steel filter, PTFE-filter or metal grid filter
Temperature rangesensor probe: according measuring rangeelectronic sensor driver device:-20...60°C >
(-4...140°F) electronic supply- and evaluation device:-40...60°C >
(-40...140°F) electronic with display:0...40°C >
(32...104°F) Storage temperature rangeelectronics and sensor head-30...60°C >
(22...140°F) Electromagnetic compatibility accordingEN61000-6-4 EN61000-6-2 >
1)Refer to the working range of the humidity sensor.3) Refer to accuracies of calculated values.2) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement). EE0E
20