TT D D EE2 EE2
Measurement valuesRelative humidity
Humidity sensor >
1) HC1000-400 Working range >
1) 0...100% RHAccuracy >
*) (including hysteresis, non-linearity and repeatability, traceable to intern. standards, administrated by NIST, PTB, BEV...)-15...40°C >
(5...104°F) <90% RH± (1.3 + 0.3%*mv) % RH-15...40°C >
(5...104°F) >90% RH± 2.3% RH-25...70°C >
(-13...158°F) ± (1.4 + 1%*mv) % RH -40...180°C >
(-40...356°F) ± (1.5 + 1.5%*mv) % RH Temperature dependence of electronicstyp. ± 0.01% RH /°C(0.0055% RH / °F) Response time with metal grid filter at 20°C / t >
90 < 15s >
Temperature
Temperature sensor manufacturer elementPt1000 (Tolerance class A, DIN EN 60751)
Working range sensing head EE29-xAx-40...60°C >
(-40...140°F) EE29-xBx-40...80°C >
(-40...176°F) EE29-xDx-40...180°C >
(-40...356°F) EE29-xEx-40...180°C >
(-40...356°F) Accuracy (typ.) >
∆ °C °C Temperature dependence of electronicstyp. ± 0.005°C/°C >
Outputs
2) Two freely selectable and scaleable analogue outputs0 - 5V-1mA< IL< 1mA0...100% RH / xx...yy°C respectively0 - 10V -1mA< IL< 1mA4 - 20mAR >
L < 500 Ohm 0 - 20mAR >
L < 500 Ohm >
General
Supply voltageSELV 8...35V DC >
SELV = Safety Extra Low Voltage SELV 12...30V AC (optional 100...240V AC, 50/60Hz)Current consumption- 2x voltage outputfor 24V DC/AC:typ. 40mA - 2x current outputtyp. 80mA Pressure range for pressure tight sensor0.01...15bar (0.15...218psi) System requirements for software WINDOWS 98 or later; serial interface Housing / protection classPC or Al Si 9 Cu 3 / IP65; >
Nema 4 Cable gland M16 x 1.5 cable Ø 4.5 - 10 mm >
(0.18 - 0.39”) Electrical connection screw terminals up to max. 1.5mm² >
(AWG 16) Sensor protectionstainless steel sintered filter, PTFE filter or metal grid filterWorking and storage temperature range of electronics-40...60°C >
(-40...140°F) -20...50°C >
(-4...122°F) - housing with displayElectromagnetic compatibility according toEN61000-6-2EN61000-6-3ICES-003 ClassB EN61326-1+A1+A2 FCC Part15 ClassB >
1) Refer to the working range of the humidity sensor. 2) Can easily be changed by software.*) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement). EE2
14