Normalization See “Standardization”. Mass Loading The change in actual sensitivity of a back-to-back
accelerometer when loaded with masses signifi-
cantly different from the transfer standard with
which it was originally calibrated. Correction curves
of actual sensitivity vs. frequency are referred to as
“mass loading correction curves”. Phase Shift The difference in phase angle between input meas-
ured and output electrical signal, measured in
degrees. The response may lag or lead the input. Picocoulomb A measure of electrostatic charge output frompiezoelectric sensors. 1 Picocoulomb= 1 x 10 Linearity Actually the non-linearity or deviation from a
straight line in a plot of output amplitude vs. input
amplitude of a sensor or system. At Dytran, we use
the popular zero based best straight line method
of determining linearity. Full scale is determined, a
zero based best straight line plot of output vs. input
is made, an error band is created using a specified
percent of full scale. All points on the curve must
fall within this error band. Triboelectric Noise Zero Shift Transient Thermal Response The unwanted generation of electrical charge by
the chafing together of insulation layers inside
industrial electrical cables during flexing caused by vibration
and shock induced motion. The change in baseline level of the output voltageof an accelerometer immediately after a mechani-cal shock. The time history of the change in output (voltage or
charge) from a sensor resulting from a sudden
change in ambient temperature. Transverse Sensitivity Thermal Coefficient of Sensitivity The output of an accelerometer in response to
motion in directions orthogonal to its sensitive axis,
expressed as a ratio of measured output to cross
axis input, in percent. The measured change in the sensitivity (of scale
factor) of a sensor, from its room temperature (ref-
erence) value to the value at a higher or lower sta-
bilized ambient temperature. This parameter is
measured in percent of change in sensitivity per
degree of temperature deviation. Standardization As applied to systems, refers to the application of
gain or attenuation as needed to correct the sensi-
tivity of a sensor to the exact nominal (or design)
sensitivity. As applied to sensors, refers to the
process by which the sensor sensitivity is brought
close to the nominal value. In specification sheets,
it is the highest acceptable deviation of the meas-
ured sensitivity from the nominal value, expressed
in percent. Sensor Bias When IEPE sensors are supplied with constant cur-
rent, a DC voltage bias exists at the center conduc-
tor of the sensor connector. The output signal is
superimposed on this DC bias. The power unit
extracts the signal from this bias by various means,
effectively “blocking” it. The normal range for this
voltage is +9 to +12 Volts. >
glossary of TERMS
Acceleration Compensation A design incorporating components within the sen-sor to cancel the effect of motion (vibration) on the
sensor output signal. Acceleration Sensitivity In a
industrial pressure transducer or impulse hammer, this
refers to the unwanted output signal from the sen-
sor in response to vibration in the sensitive axis of
the sensor. This parameter is specified in terms of
psi/g for
pressure sensors and Lbs/g for impulse
hammers. Certain Dytran sensors such as the Series
2200 and 2300 pressure sensors and the
Dynapulse LIVM Frequency Response Low Impedance Voltage Mode (LIVM=IEPE),
Dytran’s Trademark describing its line of piezoelec-
tric sensors with built-in impedance converting elec-
tronics. The highest and lowest frequencies at which meas-
ured deviations from a reference sensitivity (usual-
ly 100 Hz) lie within a specified error. The devia-
tions are usually specified as -5%, but in some
cases -3dB is specified. Fault Monitor Meter Discharge Time Constant (TC) A
Dc voltmeter incorporated into IEPE
power units
to read the DC voltage at the output of the current
source. If a sensor is connected, this meter reads
the sensor bias voltage and can be used to indicate
open and shorted cables and sensors or to verify
normal system operation. >
™ series of impulse hammers are acceler-ation compensated to minimize this effect. The time required for the output voltage from a
sensor or system to discharge to 37% of its original
value in response to a zero rise time step function
input. This parameter determines low frequency
response. Charge Mode Current Source Power Unit Sensors which contain piezoelectric
industrial crystals but no
built-in electronics. These sensors may use quartz
or
piezoceramics to generate electrostatic charge
signals in response to input measured. A power supply expressly designed for use with
Low Impedance Voltage Mode (LIVM=IEPE) sensors
consisting of an energy source (batteries of
Industrial dc power supplies) and a constant current element.
These units are characterized by stable constant
current output and high dynamic impedance. They
also serve as signal decoupling devices. >
GLOSSARY OF TERMS Rise Time Sensitivity The time required for a sensor or system to respond
to an instantaneous input step function, measured
from the 10% to 90% points on the response wave-
forms. The scale factor of a sensor or system, measured in
terms of change in output signal per change ininput measured. Accelerometer sensitivity is
expressed in terms of mV or pC/g, pressure is
expressed in terms of mV or pC/psi and force sen-
sor sensitivity is expressed in terms of mV or
pC/LbF. >
-12 Coulomb. Resonant Frequency Also called “natural frequency”. The lowest fre-
quency of a second order system (spring-mass sys-
tem) which satisfies the solution to the differential
equation of motion. At this frequency, the ampli-
tude increases by factors of as much as 100. The
usable upper frequency of a sensor is determined
by this parameter. All sensors are spring-mass sys-
tems with intrinsic natural frequencies. RMS Root Mean Square. In AC theory, this is the “heat-
ing value” of an AC waveform and it is equal in
amplitude to .707 times the peak value of the
waveform. Most AC meters are calibrated to read in
RMS. To obtain the peak value of the waveform,
multiply the RMS value by 1.414. >
33