CPG 8000
Pressure range «
bar'
0 ... 0.1 up to 0 ... 2500 (sensor related]
Total uncertainty of measurement "
0.05 % FS up to O.OOS % IS » (sensor related)
Pressure ports
- up to 400 bar (standard version)
6 mm SWAG E LO K"*' tube fitting
- over 400 bar (high pressure version)
WSNOTRIK" tube fitting
Pressure ports adapter
on request
Tube material
stainless steel 1.4571 & 1.4401 (all connections without seals)
■.....■ i...... i I'.' 1) ii:
safety valves, scaled to 120% of FS
Instrument version
desktop/19" plug-In case
Screen
TFT-Colour graphics display (320 a 240 Pixel)
Screen division
windows
1,801-4
Indication of measured values
digits
up to 7
Measuring rate
values/sec.
up to 8 [sensor related)
Keyboard
Membrane keyboard
Evaluation unit
Motorola 50 MHz Power PC-Board
Operating system
Windows CE
Digital Interface
RS-232 and IEEE-488.2
8 programmable change-over contacts
Power supply
100-240 VAC, SO / 50 Hz, <55 VA
Permissible
Media
sensor related; for any noncorrosive liquid resp. clean, dry air or nitrogen
Operating temperature
°C
15 ..35
Storage temperature
°C
0... 70
Air humidity
%
35 ... 85 relative humidity without moisture condensation
Ingress protection
IP31 (front panel IP41)
Weight
Kg
10 ... 17 (dependent on sensors used)
Dimensions
sea technical drawings
CE-mark
conformity certificate
Calibration i
with DKD calibration certificate; others on request
Wiring details
The input socket and the digital interfaces (see Rear view on Page 2) should/must only be connected with specific connection cables approved tor this purpose.
Relay outputs (8 potential free change-over contacts)
Condition of wiring scheme as shown: inactive Output capacity of relays: max. 1 A / 30 VDC
rTĖTTĖrTirfi