Portable Pressure Calibration Kits LR-Cal LPP-KIT - DRUCK & TEMPERATUR Leitenberger - #23 |
|
You may also be interested in
Capacitive sensor, Chart recorder, Hand pump, Capacitive level sensor, Recorder
Text version of the page
Electronic Pressure Calibrator LPC 300 LPC 300
multi functional, Accuracy ±0.025% FSO
DRUCK & TEMPERATUR Leitenberger GmbH
Bahnhofstr. 33 C D-72138 Kirchentellinsfurt C Germany
Tel.: +49-7121-90920-0 C Fax: +49-7121-90920-99
E-Mail: DT-Export@Leitenberger.de C http://www.druck-temperatur.de
Technische Änderungen vorbehalten. Freibleibend C (Rel. 060713) • All technical modifications reserved. Without engagement
LPC 300 electr. measurement plugs
voltage & current input, voltage output,
pressure switch, USB/RS232, battery charger
Technical Data:
Available LPC 300 reference pressure sensors
(special ranges on request)
Pressure Range
[bar]
overpressure
[bar]
burst pressure
[bar]
Accuracy: ±0.025% FSO *)
Overpressure warning: audio visual
Temperature compensation: active, 0...50°C
Wetted parts: stainless steel, welded; Thread: 1/2" BSP male
Resolution: adjustable, max. 6 digits +prefix +decimal point
Measurement voltage: 0-10 V, 0-5 V, 0-1 V
Resolution: display x 0,1 mV, accuracy ±0.5 mV
Measurement current: 0-20 mA, 4-20 mA
Resolution: display x 1 ìA, accuracy ±1.6 ìA
Voltage supply: 24 VDC min. 20 mA, max. 50 mA
tolerance ±1V
Operation conditions: 0°C... 50°C, max. 80% r.h. non-cond.
(during battery charging: 0...45°C)
Storage: -20°C...+60°C, max. 80% r.h. non-condensing
Graphic display: 128 x 128pPixel, display 44.7 x 44.7 mm
Storage capacity: 16 units under test, each 22 test points
RS232-Parameter: 4800, 9600 or 11200 Baud, adjustable
Battery: Lithium-Ion with intelligent charging electronics
Battery charger: 230 VAC 50/60 Hz (other on request)
Electrical connections:
charging plug 9V / 450 mA ±50 mA, with metal protection cap
PC communication: USB and RS232, with metal prot. cap
Measurement plugs:: 4 mm standard plugs for:
current measurement 0/4-20 mA
voltage measurement 0-1/5/10 V
pressure switch
Voltage supply: 24 V / 50 mA
Dimension: 12.5 x 21 x 8 cm (width x height x depth)
Weight: appr. 1.1 kg
0...0.25 gauge 1.6 2.4
0...0.4 gauge or absolute 2 2.4
0...0.6 gauge or absolute 4 4.8
0...1 gauge or absolute 5 6
0...1.6 gauge or absolute 10 12
0...2.5 gauge or absolute 10 12
0...4 gauge or absolute 17 20.5
0...6 gauge or absolute 35 42
0...10 gauge or absolute 35 42
0...16 gauge or absolute 80 96
0...25 gauge 50 96
0...40 gauge 80 400
0...60 gauge 120 550
0...100 gauge 200 800
0...160 gauge 320 1000
0...250 gauge 500 1200
0...400 gauge 800 1500
0...600 gauge 1200 1500
0...1000 gauge 1500 3000
-0.4...0 gauge 2 2.4
-0.6...0 gauge 4 4.8
-1...0 gauge 5 6
-0.25...+0.25 gauge 1.6 2.4
-0.4...+0.4 gauge 2 2.4
-0.6...+0.6 gauge 4 4.8
-1...+1.5 gauge 10 12
-1...+3 gauge 17 20.5
-1...+5 gauge 35 42
-1...+9 gauge 35 42
-1...+15 gauge 80 96
-1...+24 gauge 50 96
*) calibrated at +23°C, incl. linearity, hysteresis and repeatability
The LPC 300 can manage up to 5 system-sensors
at a time.
Optional Accessories:
• rubber protection caps for electrical measuring plugs
• electr. extension cable for reference sensor
(herewith the reference sensor can be used “outside”
the LPC 300 unit for more flexibility)
|
|