catalogue search
|
You may also be interested in
Filter, Detector, Suit, Infrared camera, Gas sensor
Text version of the page
|299Hydrogen Sulfide 1/dfor use in Multi-test Med. GasesOrder No.81 01 831Application Range Standard Measuring Range:1 to 20 ppm Test Volume: 1 L Flow Rate: 0.17 L / min (CO2 )Time for Measurement:6 minStandard Deviation: ± 15 % Dräger sampling tubes and systems - 20351 HColor Change: white brown Ambient Operating Conditions Temperature: 2 to 40 °CAbsolute Humidity:max 40 mg H2 O / LPressure:The tube may only be used for depressurized compressed air Reaction Principle H 2 S + Cu 2+ CuS + 2 H + Cross Sensitivity 500 ppm hydrochloric acid, 500 ppm sulfur dioxide, 500 ppm ammonia or 100 ppm arsine do not affect the indication.Methyl mercaptan and ethyl mercaptan change the entire indicating layer to a pale yellow. when mixed with hydrogen sulfide the reading is extended by approx. 30 %. Evaluation reading on the (n= 10) scale = ppm H 2 S ST-131-2001 |