SPECIFICATIONS
Pressure Pa: ± 0 to 99.9 ± 100 to 999 ± 1.00 to 7.50 Kpa mmH >
2 O: ± 0 to 0.999 ± 1.00 to 9.99 ± 10.0 to 99.9 ± 100 to 750inH >
2 O: ± 0 to 0.999 ± 1.00 to 9.99 ± 10.0 to 30.0 mbar: ± 0 to 0.999 ± 1.00 to 9.99 ± 10.0 to 75.0 Velocity m/sec: 0.8 to 9.99 10.0 to 99.9 ft/min: 161 to 20,000 Accuracy:@ 20°C Readings < 100 counts ± 2 counts Pressure and Velocity Readings > 100 counts ± 1% of reading ± 1 count Recommended operational limits: 0° to 50°C (32° to 123°F) Span Stability v. Temperature: Better than 0.1% of range in use per °C (per 2°F) Zero Drift: Negligible due to autozero system. When auto zero set at 1 minute intervals. (2 minute warm up) Zero System Accuracy: ± 1 count (any range) Orientation Effect: (any 45° change) 0.1 pascal typical Output Socket: RS 232 (baud rate 9600)
Data logging: Up to 2500 any units Software: Down load readings to PC in a very basic form. (Optional)
Power source: Dry cell (MN1604, PP3) Rechargeable 8.4v 120 mAh System Air Leak (Typical): 0.1ml/minute at 5Kpa Safe Line / Differential Pressure: 15 Kpa Storage Temperature Limits: -5° to + 50°C Dimensions: 45 x 92 x 185 mm Weight: 555 grammes with battery >
DP MEASUREMENT Unit 11, Top Angel, Buckingham Industrial Park, Buckingham. MK18 1TH. Tel/Fax ++44 (0) 1280 823823 www.ttseries.com In the interest of continuous product development and improvement, DP Measurement reserves the right to amend the specifications, models, features and colour of the TT Series Micromanometers at any time without prior notice.