20 GLP 22 (continuation)
Technical specifications
Operation
The GLP 22 software has been designed so that the operator can usethis high-performance pH-meter by following on-screen instructions, using a simple keypad and practically without having to read the instructions manual.
Measuring ranges
pH -2 ...16,mV± 2000, °C -20 ...150 and relative mV, mol /I10-5 ...10"1 (or ppm org/l).
Resolution
Standard, 0.01 pH, 1 mV,0.1°C. Selectable, 0.1, 0.001 pH, 0.1 mV.
Measuring error (± 1 digit)
s 0.003 phU0.2 mV.sO.2 °C.
Reproducibility (± 1 digit)
±0.01 pH,±1 mV,±0.1°C. ±0.001 pH,±0.1 mV,±0.1 °C.
Input impedance
>1012Ohm(n).
Autom. temperature compensation
With Pt 1000ATC probe or entering data using the keypad.
pH calibration
ln2or3 points, validity 0h...7 days.
Special calibration, 1 selectable point between 0...14 pH.
Recalibration warning with option of forced recalibration.
Rejection of defective electrodes.
Accepted electrode values:
-slope 50... 65 mV/pH,at25°C.
- sensitivity 86... 110%.
- asymmetry potential ± 40 mV.
Autom. buffer recognition
Technical pH 2.00,4.01, 7.00, 9.21,10.90, (at25°C). NIST pH 1.679,4.0061, 6.865, 9.180,10.254, (at 25°C).
History of pH electrode
Last 10 calibrations. No. of measurements, etc.
Measuring programs
A fixed one and four user-modifiable ones.
Data logger
Storage capacity: 447 readings. Alphanumeric and graphical reports
Display
Alphanumeric backlit crystal display. 4 lines of 20 characters.
Languages
English, Spanish, french, italian.
Power up diagnostic
RAM, EPROM, EEPROM memories, display and keypad.
Inputs and outputs
Measuring electrode, BNC connector.
Reference electrode, banana 0 4.
ATC, Pt 1000 type, telephone or banana connector 0 4.
Interface for standard PC keyboard.
Interface RS 232 C, bidirectional.
Polarising current, 10uA (i.e. for Karl Fischer).
Analog outlet, electrode potential follower.
Control for CRISON stirrer: On/Off and speed.
Power supply
12VDC/275 mA external power supply
Electrical safety
Meets EC, EN 61010.
EMC (Electromagnetic compatibility)
MeetsEC, EN 50081-1 andEN 50082-1.
Permitted ambient conditions
Operatingtemperature 5...40°C
Storage and transport temperature -15... 65 °C.
< 80 % relative humidity, non condensing.
Enclosure
Casing material, high density polyurethane.
Physical parameters
Weight 1.580 g, dimensions 180x215x 130 mm.
pH
Communications software GLP-PC
Software forWindows, making capture and availability of data on PC easier.
The GLP 22 can be operated from a PC thanks to its two-way RS 232 C interface.
* pH 7,01
.... v. I. . - I - r . .' - . J. -» .
- - -~f - - r - ~ f "~" " '. ' ~ ~* ~~ -~ ' ~ -
' — -V-/' - - -
j1
PC display example
Specifications suject to change without notice
Ordering information
Code
22 00
GLP 22 with accessories, without electrode
22 01
GLP 22 with accessories and electrode 52 01, with integrated ATC
22 02
GLP 22 with accessories and electrode 52 02
22 03
GLP 22 with accessories, electrode 52 02 and ATC 55 31
22 04
GLP 22 with accessories, electrode 52 01 and magneticstirrer
Certificates
The GLP 22 is supplied with the specifications Certificate and the EC Declaration of Conformity as per directive EMC 89/336/EC. For further information, please see page 132.
PLUG-IN POWER SUPPLY: The instrument is supplied with a 230 VAC plug-in power supply. For 115 VAC, please specify when ordering.
SENSORS: Electrodes. Pleaseseepages36... ATC Temperature probe, Pleaseseepage44 ACCESSORIES: flexible arm, electrode cable and 250 ml bottles of pH 4 and pH 7 buffer and electrolyte. OPTIONAL ACCESSORIES: See page 61
Laboratory Catalogue 2004 1st Edition