Fluorine Determination in the Analysis Sample of Solid Waste
2Pages

{{requestButtons}}

Catalog excerpts

Fluorine Determination in the Analysis Sample of Solid Waste - 1

Fluorine Determination in the Analysis Sample of Solid Waste Sample Preparation Samples may be analyzed as air dried Crush and grind to make the sample size <0.2mm Instrument 5E-FL2350 Fluorine/Chlorine Analyzer Instrument Settings Oxygen : Flow 500ml/min; purity 99.5% Vapour :2ml/min Main Furnace Temperature:1100 °C Condensator temp. :20°C Calibration Standard: GBT4633-2014 Nominal Sample Mass: 100 mg Analysis Time: 45 minutes/per test Special Modifications to Unit None Method Approval GBT4633-2014 and ISO11724 and ASTM D5987

Open the catalog to page 1
Fluorine Determination in the Analysis Sample of Solid Waste - 2

1. Prepare instrument as outlined in the operator' s instruction manual (check the solutions, pure water content and electrodes, oxygen gas flow and etc.) 2. Do three blanks sample "Fkt" titration (initialize the titration system) 3. Combustion pretreatment with two empty crucibles (this process only for first time of using of each day) 4. Weigh a portion (0.5g) of solid waste sample in crucible, then cover the sample with a flat layer of silica sand 5. Load two crucibles with sample as prepared in step4 , then" start" analysis 6. 2 samples of calibration substance to check the calibration...

Open the catalog to page 2