Ash Fusibility Determination for Biomass Sample Preparation Crush & Grind the sample to size ≤ 1mm, insert the sample into furnace under 815℃ to burn completely ,then grind the ash to the size ≤ 0.1mm. Instrument 5E-AF4115/AF4105/AF4000 Ash Fusion Determinator Instrument Settings Choose the analyzing atmosphere, standard and how many ash sample you want to analyze. Special Modifications to Unit None Method Approval UNI CEN/TS15370 GB/T 30726 Procedure 1. Prepare the test pieces by using the mould according to different standard 2. Mount the test pieces to the support 3. Insert the test pieces into furnace under the controlling of software 4. Choose how many sample you want to analyze 5. Start analyzing to get the result Note: Do Calibrating refer to operator’s instruction manual for details. Website: www.ckic.net E-mail: info@cki