Technical data
Rondcom 41Rondcom 44/54 Measuring range Max. measuring diameter >
250 mm300 mm
Max. workpiece diameter >
400 mm580 mm
Max. load capacity >
25 kg
30 kg Rotational axis (C axis) Faceplate diameter >
148 mm
220 mm Radial accuracy 0.040 + 6H/10,000 >
µm (bandwidth)
0.020 + 4H/10,000 >
µm
0.020 + 3H/10,000 >
µm (tolerance)
Axial runout 0.020 + 4R/10,000 >
µm
Measuring speed 6 u/min2–10 u/min Centering range >
± 2 mm± 2 mm
Leveling range >
± 1°± 1°
Automatic Alignment >
NoRondcom 44 no/Rondcom 54 yes
Automatic speed adjustment >
6/10/20 u/min
Resolution >
0.1°0.025°
Vertical axis (Z axis) Measuring path >
300 mm300 mm (500 mm)
Straightness accuracy 0.5 >
µm/100 mm, 1.5 µm/300 mm
0.12 >
µm/100 mm, 0.2 µm/300 mm
Parallelism deviation from rotational axis 3 >
µm/300 mm
0.8 >
µm/300 mm
Measuring speed >
0.6–6 mm/s0.5–6 mm/s
Travel speed >
max. 15 mm/smax. 50 mm/s
Horizontal axis (R axis) Measuring path >
125 mm170 mm
Straightness accuracy 0.8 >
µm/150 mm
Linear scale >
NoYes
Parallelism deviation from rotational axis 1 >
µm/150 mm
Measuring speed >
0.5–6 mm/s
Travel speed >
5 mm/smax. 25 mm/s
Accuracy of the linear scale >
(2+L/170) µm
detector-and-stylus system CNC
swivel probe system >
NoR 54 optional
Measuring force >
30–100 mN (adjustable)30–100 mN (adjustable)
Collision protection >
Mechanical/electronicalMechanical/electronical
Measuring range max. >
±
1000 >
µm (adjustable)
max. >
±
1000 >
µm (adjustable)
Resolution max.0.001 >
µm
max.0.001 >
µm
Stylus, standard L=54 mm, D=1.6 mm hard metalL=54 mm, D=1.6 mm hard metal Other Information Power supply >
100–240 V AC 50/60 Hz100–240 V AC 50/60 Hz
Power consumption >
approx. 600 VA approx. 600 VA
Compressed air supply >
0.3–0.7 MPa0.3–0.7 MPa
Total weight >
approx. 120 kgapprox. 170 kg
Rondcom 72Rondcom 75 Measuring range Max. measuring diameter >
2–450 mm2–450 mm
Max. workpiece size >
X=600 mm/Y=550 mm/Z=760 mmX=800 mm/Y=680 mm/Z=760 mm
Max. load capacity >
200 kg200 kg (1000 kg)
X measuring range >
600 mm700 mm (1200 mm)
Y measuring range >
50 mm200 mm
Z measuring range >
1000 mm1000 mm (1500 mm, 2000 mm)
Accuracy tolerances Radial runout tolerance 0.03 >
µm standard detector-and-stylus system L=235 mm
0.1 >
µm standard detector-and-stylus system L=76 mm
0.3 >
µm detector-and-stylus system L=620 mm
0.2 >
µm detector-and-stylus system L=700 mm
Axial runout tolerance 0.1 >
µm/R 50 mm standard detector-and-stylus system
0.1 >
µm/R 50 mm standard detector-and-stylus systemL=235 mmL=76 mm
0.2 >
µm/R 50 mm detector-and-stylus systemL=620 mm
0.2 >
µm/R 50 mm detector-and-stylus system L=700 mm
Straightness 2 >
µm/200 mm standard detector-and-stylus system
0.3 >
µm/50 mm, 0.5 µm/100 mm L=235 mmstandard detector-and-stylus system L=76 mm
5 >
µm/600 mm detector-and-stylus system L = 620 mm
1.5 >
µm/700 mm detector-and-stylus system L=700mm
Parallelism 2 >
µm/100 mm
1.5 >
µm/100 mm
Positioning accuracy >
± 50 µm± 50 µm
Speed Rotation speed >
2 u/min (measuring), 10 u/min (centering)2 u/min (measuring), 10 u/min (centering)
Linear speed >
0.6 mm/s–6 mm/s 0.6 mm/s–6 mm/s
Detector-and-stylus system Stylus tip >
D=1.6 mm hard metal standardD = 0.5 mm sapphire
Stylus lenght >
Standard L=24 mm (1:1)Type A L=76 mm (1:1) Type B L=201 mm (2:1)
Measuring force >
0.07 N StandardType A 0.17 N Type B 0.085 N
12