Catalogue Rapidox 4100 Oxygen Gas Analyzer
www.cambridge-sensotec.co.uk
print switch display
Page / 1
Make a request   
Cambridge Sensotec - 147569, 142190
/ 1
See other catalogues for Cambridge Sensotec
Text version of the page
Rapidox 41 00 Portable Oxygen Gas analyser
Description
The Rapidox 41 00 electrochem ical cell oxygen gas analyser based on the Hitech G1 01 0 OEM m odu le is h ou sed in ou r sm art Rapidox m etal case with carry han dle an d tiltin g feet. Incorporating a powerful and silent linear piston pum p for remote sampling of gas via 6/4m m flexible tu bin g to the fron t pan el in pu t an d ou tput u n ion s. A flow gau ge allows th e u ser to control the rate of gas flow into th e analyser. Use galvan ic cell pri­mary elements and can measure oxygen concentrations from 1 00% to 1 00ppm in a variety of gases. The instruments are micro-processor-controlled and provide excep­tional performance and versatility. Two alarm channels, user-programmable to be high, low or off, each provide one set of volt-free changeover contacts, and can be set to any concentration within the span of the instrument. A user program mable ana­logue output can be set to one of several ranges. A choice of 0 to 5v or 4 to 20mA is provided at the tim e of ordering. Several versions of sensor are available, enabling the analyser to operate at optimum performance in a diverse range of applications. These include a sensor suitable for operating in the presence of high concentrations of mildly acidic gases such as carbon dioxide and hydrogen sulphide. See the chart below for the option s available:
Applications
• Air separation plants & Purge gas monitoring
• Nitrogen purged soldering system s
• Cylinder gas quality
• Food packaging and storing
• Landfill gas monitoring
• Glov e b ox es
• Workplace air monitoring
Features
Exceptional dynamic Range. Measurement range 100ppm to 100% oxygen. Microprocessor controlled.
Programmable alarm channels and analogue outputs (0-5V or 4-20mA). Easy to calibrate by the user using single point gas. Multi-lit auto-ranging LCD display
Secveral sensor types available for optimum performance Powerful internal linear piston pum p fitted • Full sample control with panel mounted flow gauge
Technical Data
Voltage
110 / 220V ac 50/60 Hz
Analyser dimension
205mm X 160mm X 150mm
Weight
3 kg
Display
Multi-digit LCD - character height 12.7mm
Display Ranges
E version - display range 0. 01 % to 1 00. 0% . Suitable for sam -ples con tain in g m ildly acidic gases; eg, carbon dioxide, h ydro-gen sulphide etc N version - display range 100ppm to 50%
Display Resolution
0.1% from 10% to 100% 0.01% from 0.50% to 9.99% 10ppm from 500ppm to 4999ppm 1 ppm from 50ppm to 499ppm
Stability
Better th an 2% of fu ll-scale per m on th
Cell Life
E version - up to 5 years N versions - up to 2 years
Sample Flow
Between 1 00 and 500ml/min for optimum operation. Max 1 litre/mm.
Technical Data
Working Temperature
-5 to +40°C (non condensing)
Sample Connections
In let an d ou tlet: captive-seal com pression fittin gs suita­ble for 0. 25 inch (or 6mm) o/d tube
Speed of Response
T90 - variable depending on sensor and concentration. Approximately 3s at % levels and 20s at ppm levels. The ppm figure assum es that the sensor is purged down to the ppm concentration of interest.
Analogue Outputs
0 to 5V or 4 to 20mA (specify at time of order) Each u ser-program m able to th e followin g spans:
E version 0 to 100% , 0 to 25% , 0 to 5%
N version 0 to 100% , 0 to 25% , 0 to 5% , 0 to
5000ppm, 0 to 500ppm
Alarm Outputs
Two alarms, each user-programmable for mode - HIGH, LOW or OFF Hysteresis programmable from 0 to 1 0% of set point Volt-free changeover contacts rated at 250V r. m. s. a. c m ax. , 3A m ax. , m ax power 60W, normally en­ergised.
Cambridge Sensotec Limited
31 Elizabeth Court
St Ives
CAMBS
PE27 5BQ
England
Tel: + 44 (0)1480 385812 Fax: +44 (0)1480 352529 Mob: +44 (0)7866 624236
Sa le s@ca m b r i d g e -se n sote c.co.u k Www.cam bridge-sensotec.co.uk
DirectIndustry's Virtual Technical Library: PDF Catalogue | Technical Documentation | Brochure | Manual | Industrial directory | Specifications | Characteristics
Search Go
page 1 p.1
pdf-page pdf di En 2008-10-41-12