Fusion FaroArm
Performance Specifications
Model
Single Point Articulation
Volumetric
FaroArm
(Measuring Range)
Performance Test (Max-Min)/2
Maximum Deviation
Weight
axis
6 7
6 7
6 7
Fusion
±.036 mm
±.046 mm
±.051 mm
±.064 mm
9.30 kg
9.50 kg
1.8 m (6 ft. )
(±.0014 in.)
(±.0018 in.)
(±.0020 in.)
(±.0025 in.)
(20.5 lbs.)
(21 lbs.)
Fusion
±.043 mm
±.051 mm
±.061 mm
±.086 mm
9.50 kg
9.75 kg
2.4 m (8 ft. )
±.0017 in.
(±.0020 in.)
(±.0024 in.)
(±.0028 in.)
(21.0 lbs.)
(21.5 lbs.)
Fusion
±.074 mm
±.089 mm
±.104 mm
±.124 mm
9.75 kg
9.98 kg
3.0 m (10 ft.)
(±.0029 in.)
(±.0035 in.)
(±.0041 in.)
(±.0049 in.)
(21.5 lbs.)
(22 lbs.)
Fusion
±.104 mm
±.124 mm
±.147 mm
±.175 mm
9.98 kg
10.21 kg
3.7 m (12 ft.)
(±.0041 in.)
(±.0049 in.)
(±.0058 in.)
(±.0069 in.)
(22.0 lbs.)
(22.5 lbs.)
FaroArm Test Methods - (Test methods are a subset of those given in the B89.4.22 standard.)
Single Point Articulation Performance Test (Max-Min)/2: The probe of the FaroArm is placed within a conical socket, and individual points are measured from multiple approach directions. Each individual point measurement is analyzed as a range of deviations in X, Y, Z. This test is a method for determining articulating measurement machine repeatability. Volumetric Maximum Deviation: Determined by using traceable length artifacts, which are measured at various locations and orientations throughout the working volume of the FaroArm. This test is a method for determining articulating measurement machine accuracy.
Hardware Specifications
Operating Temp range: 10°C to 40°C (50°F to 104°F) Temperature Delta: 3°C/5min. (5.4°F/5min.)
Certifications: MET (UL, CSA Certified) CE compliance
Directive 93/68/EEC, (CE Marking)
Directive 89/336/EEC, (EMC)
FDA CDRH, Subchapter J of 21 CFR 1040.10
Electrical Equipment for Measurement, Control & Lab Use
EN 61010-1:2001, IEC 60825-1, EN 61326
Electromagnetic Compatibility (EMC)
EN 55011, EN 61000-3-2, EN 61000-3-3, EN 61000-4-4,
EN 61000-4-5, EN 61000-4-6, EN 61000-4-8, EN 61000-4-11
HumicHty:
Calibration Lifecycle: Power Supply:
95%, noncondmsmg Permanent
Universal worldwide voltage
85-245VAC, 50/60 Hz
r^^Hirtificd ISO by
Efoi
ISO-17025 : 2005
ACCREDITED
Certificate # L1147
LABORATORY
ACCREDITATION
BUREAU
Global Sales Offices: USA • Germany • Swiss • France • United Kingdom • Spain • Italy • Netherlands • Poland • Singapore • China • Japan • India • Brazil
freecall: 00800 3276 7253
www.faro.com
Revised: 29 October 2007
© 2007 FARO EUROPE GmbH & Co. KG
x:\control\referenc\04market\english\04ref201-268.pdf
DirectIndustry's Virtual Technical Library: PDF Catalogue | Technical Documentation | Brochure | Manual | Industrial directory | Specifications | Characteristics
Search Go
page 1 p.1
page 2 p.2
pdf-page pdf di En 2008-11-47-22