| | | FaroArm Test Methods - (Test methods are a subset of those given in the B89.4.22 standard.) Single Point Articulation Performance Test (Max-Min)/2: The probe of the FaroArm is placed within a conical socket, and individual points are measured from multiple approach directions. Each individual point measurement is analyzed as a range of deviations in X, Y, Z. This test is a method for determining articulating measurement machine repeatability. Volumetric Maximum Deviation: Determined by using traceable length artifacts, which are measured at various locations and orientations throughout the working volume of the FaroArm. This test is a method for determining articulating measurement machine accuracy. | | |
| | | Operating Temp range: 10°C to 40°C (50°F to 104°F) Temperature Delta: 3°C/5min. (5.4°F/5min.) | | Certifications: MET (UL, CSA Certified) CE compliance Directive 93/68/EEC, (CE Marking) Directive 89/336/EEC, (EMC) FDA CDRH, Subchapter J of 21 CFR 1040.10 Electrical Equipment for Measurement, Control & Lab Use EN 61010-1:2001, IEC 60825-1, EN 61326 Electromagnetic Compatibility (EMC) EN 55011, EN 61000-3-2, EN 61000-3-3, EN 61000-4-4, EN 61000-4-5, EN 61000-4-6, EN 61000-4-8, EN 61000-4-11 | | |