Fyrite
®
Fast, accurate and easy to use instruments for measuring andanalyzing carbon dioxide or oxygen. Fyrite
Analyzers specification are available
for either CO >
2 or O >
2 analysis, and each model is produced in threescale ranges. All six instruments are similar in appearance and size, but differ inimportant construction details, as well as in the absorbing fluids.Each model, therefore, is suitable only for the particular gas analysis or scale range for which it has been manufactured.
Accuracy is within ±1/2% CO >
2 or 0 >
2 . >
Operation Applications
Fyrite absorbing fluid is selective in the chemical absorption
of carbon dioxide or oxygen, respectively. Therefore, the
Fyrite’s accuracy, which is well within the range required
for industrial and professional applications, does not
depend upon complicated sequential test procedures. In addition, Fyrite readings are unaffected by the presence
of most background gases in the sample. The number of tests possible with one fluid chargedepends on the concentration of samples being tested.
At midpoint scale reading the CO 0-7.6% CO >
2 –CO >
2 tests of controlled atmospheres in fruit, vegetable,meat storage rooms, and
professional incubator monitoring.0-7.6% O >
2 –Oxygen determination in flammable gases; oxygen tests
to check inertness of atmosphere in silos, fuel tanks, etc.0-20% CO >
2 –Flue gas combustion tests; CO >
2 tests of heat treatingatmospheres.0-21% O >
2 fluid is good forapproximately 300 gas samples and the O >
2 fluid for 100tests. The need to replace fluid can be easily determined
with a simple test, and replacement is an easy procedure.
These test procedures, as well as other good information,
are provided in the Fyrite manual 11-9026. >
2 –Flue gas combustion tests, oxygen deficiency test.
Checking oxygen concentrations in hydrogen cooled generators and oil sealed inert gas
industrial transformers.0-60% CO >
2 1.800.736.4666BACHARACH –Checking CO >
2 in inert gas blankets in tankers and bargescarrying gasoline and other combustibles; CO tests onlime kilns; checking CO >
Features
2 2 in sewage plant digesters.0-60% O Fyrite Indicators have a broad range; they may be exposed
to ambient temperatures from -30°to 150°F, and gases
up to 850°F may be tested with standard aspirator sampling equipment (special sampling equipment for
higher gas temperatures or dry gases is available). Order
Fyrite Instruction Manual 11-9026. For temperatures
above 1400°F, a ceramic sampling tube (Bacharach Part #11-0164) is available. >
2 –Oxygen test in connection with oxygen and gas anesthesiology. >
Note: United States and Foreign Postal Regulations prohibit Fyrite fluid, in or out of any unit, from being shipped parcel post. 17