100-30
table.main {}
tr.row {}
td.cell {}
div.block {}
div.paragraph {}
.font0 { font:4.00pt "Arial", sans-serif; }
.font1 { font:5.00pt "Arial", sans-serif; }
.font2 { font:7.00pt "Arial", sans-serif; }
.font3 { font:8.00pt "Arial", sans-serif; }
.font4 { font:12.00pt "Arial", sans-serif; }
.font5 { font:19.00pt "Arial", sans-serif; }
.font6 { font:20.00pt "Arial", sans-serif; }
.font7 { font:5.00pt "Franklin Gothic Medium", sans-serif; }
.font8 { font:8.00pt "Franklin Gothic Medium", sans-serif; }
.font9 { font:6.00pt "Impact", sans-serif; }
Test Bench® High Performance DMM
High Quolity,
Ruggedized,
Multifunction
391A
390A
389A
388B
High performance and value priced, the Test Bench® Séries offers more features for the dollar than other multimeters. These meters include Component Test capabilities, Resistance, Diode Test and Capacitance, in addition to measuring Frequency, Temperature and a Logic Indicator. See the chart below for the meter that best fits your needs. Dual injection molding process allows a better grip and protection for the meters. CE marked and UL listed.
Feotures
i o a et
i o/\ et
700A
1 O O D
391H
390H
389H
388B
Basic Functions
True RMS
V
Ranging
Manual
Auto/Manual
Auto/Manual
Manual
DCV Accuracy
0.05%o
0.1%
0.25%
0.5%
AC/DC Voltage and Current
V
V
V
V
Display Digits, Count
4 1/2, 20000
3 3/4, 4000
3 3/4, 4000
3 3/4, 4000
Bar Graph (41 Segment)
V
V
Capacitance Measurement
V
V
V
Transistor Test
V
Temperature Probe
V
Logic Probe
V
V
Relative Mode
V
V
Min/Max Hold
V
V
Peak Hold
V
V
Data Hold
V
V
V
RS-232
V
Common Feotures
■ Résistance measurement
■ Diode check
■ Frequency measurement
■ Audible continuity
■ All current ranges fused
■ Ruggedized case
■ Auto power off
■ Designed to meet IEC61010-1 CATIII 1000V, class 2
Specificotions
1
models
391A
390A
389H
388B
Volts
True RMS reading
DCRanges
200mV, 2V, 20V, 200V, 1,000V
400mv, 4V 40V 400V 1000V
AC Ranges
200mV 2V 20V 200V 750V
400mv, 4V 40V 400V 750V
Resolution
10/iV 100mV 1 mV 10mV 100mV
100/iV 1mV 10mV 100mV 1V
DC Accuracy
±(0.05% + 3 dgt)
±(0.1% rdg + 2 dgt)
±(0.25% rdg + 2 dgt)
±(0.5% rdg + 1 dgt)
AC Accuracy
±(1% +10 dgt.) 50Hz-500Hz
400mV: ±(1.2% rdg+ 5 dgt)
±(1.2% rdg+ 4 dgt)
±(2% + 10 dgt.) 500Hz - 2kHz,
50Hz-100Hz
750V: ±(2% rdg+ 4 dgt)
500Hz on 750Vrange
±(1.0% rdg + 3 dgt) 50Hz-500Hz
>10% of range at 200mV range
±(1.5% rdg + 3 dgt) 500Hz-1kHz
Overload Protection
1200VDC or AC rms
1100VDC or AC rms
1200VDC or AC rms
500VDC/AC rms 15 sec
500VDC/AC rms 15 sec
on 200mV range
on 200mV range
Input Impedance
400mV: >100M£2, 4V: 10M£2,
40V - 1000V: 9.1MÎ2